VENTED AUTOFEED CHAMBER
Report
- Report Number
- 9611451-2013-00418
- Event Type
- Malfunction
- Date Received
- May 24, 2013
- Date of Event
- April 12, 2013
- Report Date
- April 25, 2013
- Manufacturer
- FISHER & PAYKEL HEALTHCARE LTD
- Product Code
- BTT
- PMA / PMN Number
- K934140
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER
Narratives
(B)(4). METHOD: THE COMPLAINT MR290 VENTED AUTOFEED HUMIDIFICATION CHAMBER WAS RETURNED TO FISHER & PAYKEL HEALTHCARE IN (B)(4) AND WAS VISUALLY INSPECTED FOR THE REPORTED DAMAGE. RESULTS: VISUAL INSPECTION REVEALED THAT THE SPIKE AND THE FEEDSET TUBING WERE SEPARATED FROM EACH OTHER. A SUFFICIENT AMOUNT OF GLUE WAS PRESENT AROUND THE SPIKE TUBING CONNECTION; HOWEVER THE GLUE HAD NOT BONDED PROPERLY. A LOT CHECK REVEALED NO OTHER COMPLAINTS OF THIS NATURE FOR LOT NUMBER 120630. CONCLUSION: THE SEPARATION OF THE SPIKE AND FEEDSET TUBE WAS DUE TO THE FAILURE OF THE GLUE BOND. WE WERE UNABLE TO DETERMINE THE CAUSE FOR THIS. ALL CHAMBERS ARE PRESSURE TESTED BEFORE THEY LEAVE THE PRODUCTION LINE AND ANY HOLES OR LEAKS IN THE FEEDSET ARE IDENTIFIED DURING THIS PROCESS. THIS SUGGESTS THAT THE FAILURE OF THE BOND DEVELOPED AFTER THE CHAMBER WAS RELEASED FOR DISTRIBUTION. THE USER INSTRUCTIONS WHICH ACCOMPANY THE MR290 CHAMBER STATE THE FOLLOWING: "SET APPROPRIATE VENTILATOR ALARM." "PERFORM A PRESSURE AND LEAK TEST ON THE BREATHING SYSTEM AND CHECK FOR OCCLUSIONS BEFORE CONNECTING TO A PATIENT." (B)(4).
A HOSPITAL IN (B)(6) REPORTED VIA OUR DISTRIBUTOR THAT THE WATER FEEDSET TUBE ON AN MR290 AUTOFEED HUMIDIFICATION CHAMBER WAS LEAKING DURING SET UP, PRIOR TO PATIENT USE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 231898 | VENTED AUTOFEED CHAMBER | AUTOFEED CHAMBER | BTT | FISHER & PAYKEL HEALTHCARE LTD | MR290V | 120630 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |