FDA Adverse Event Malfunction Summary report: N

HUDSON ADULT VENTILATOR TUBING/CIRCUIT

MDR report key: 3127302 · Received May 20, 2013

Report

Report Number
3004365956-2013-00180
Event Type
Malfunction
Date Received
May 20, 2013
Date of Event
April 24, 2013
Report Date
May 8, 2013
Manufacturer
TELEFLEX MEDICAL
Product Code
BZO
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
LA, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

A VISUAL, DIMENSIONAL AND FUNCTIONAL INSPECTION OF THE PRODUCT INVOLVED IN THE COMPLAINT COULD NOT BE CONDUCTED SINCE THE PRODUCT WAS NOT RETURNED. THE DEVICE HISTORY RECORD (DHR) FOR THE REPORTED LOT NUMBER (02K1201913) WAS REVIEWED AND NO ISSUES OR DISCREPANCIES WERE FOUND WHICH COULD POTENTIALLY RELATE TO THIS COMPLAINT. THE DHR SHOWS THAT THE PRODUCT WAS ASSEMBLED AND INSPECTED ACCORDING TO OUR SPECIFICATIONS. NO NON CONFORMANCE REPORTS WERE ORIGINATED FOR THE REPORTED LOT NUMBER THAT CAN BE ASSOCIATED TO THE COMPLAINT REPORTED. FOR CAT # 1619 THE ASSEMBLY PROCESS AND MFG PROCEDURE WERE REVIEWED AND NO FINDINGS THAT CAN POTENTIALLY RELATE TO THE REPORTED ISSUE WERE FOUND. THE CUSTOMER COMPLAINT WAS NOT CONFIRMED DUE TO THE LACK OF THE PRODUCT SAMPLE TO PERFORM A PROPER INVESTIGATION AND DETERMINE THE ROOT CAUSE. THE COMPONENT INVOLVED P/N MP-0498 IS A MOLDING COMPONENT. ALTHOUGH THE CUSTOMER COMPLAINT WAS NOT CONFIRMED, BASED ON SIMILAR COMPLAINTS, THE MOLDING DEPT IS GOING TO IMPROVE THE PROCESS (PARAMETERS SHEET) IN ORDER TO ASSURE GOOD RETENTION BETWEEN THE PLUG AND TEMP PROBE PORT.

Description of Event or Problem · 1

THE COMPLAINT REPORTED AS: THE CUSTOMER ALLEGES THAT THE TEMP PROBE CAP, ON THE CIRCUIT, WILL NOT STAY CLOSED WHEN VENTILATING. THIS ISSUE DETECTED PRIOR TO PT USE DURING FUNCTIONAL TESTING/INSPECTION. NO REPORT OF PT INJURY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
223352 HUDSON ADULT VENTILATOR TUBING/CIRCUIT VENTILATOR CIRCUIT BZO TELEFLEX MEDICAL 02K1201913

Patients

Seq Age Sex Outcome Treatment
1