VITROS CHEMISTRY PRODUCTS BENZ REAGENT
Report
- Report Number
- 1319808-2013-00031
- Event Type
- Malfunction
- Date Received
- May 22, 2013
- Date of Event
- April 18, 2013
- Report Date
- May 22, 2013
- Manufacturer
- ORTHO-CLINICAL DIAGNOSTICS
- Product Code
- JXM
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- OTHER
Narratives
THE INVESTIGATION DETERMINED THAT A NEGATIVE VITROS BENZ RESULT WAS OBTAINED FROM A SINGLE PATIENT SAMPLE USING THE VITROS 5,1 FS CHEMISTRY SYSTEM. THE INTENDED USE SECTION OF THE VITROS BENZ IFU STATES THAT THE VITROS CHEMISTRY PRODUCT BENZ ASSAY IS INTENDED FOR USE BY PROFESSIONAL LABORATORY PERSONNEL. IT PROVIDES ONLY A PRELIMINARY TEST RESULT. A MORE SPECIFIC ALTERNATIVE CHEMICAL METHOD MUST BE USED TO CONFIRM A RESULT OBTAINED WITH THE VITROS BENZ ASSAY. GAS CHROMATOGRAPHY/MASS SPECTROMETRY (GC/MS) IS THE PREFERRED CONFIRMATORY METHOD. CLINICAL CONSIDERATION AND PROFESSIONAL JUDGMENT SHOULD BE APPLIED TO ANY DRUG-OF-ABUSE TEST RESULT. THE MOST LIKELY CAUSE IS A KNOWN LIMITATION OF THE VITROS BENZ REAGENT RELATED TO LOW CROSS REACTIVITY WITH LORAZEPAM. THERE IS NO EVIDENCE THAT A REAGENT RELATED ISSUE CONTRIBUTED TO THE EVENT.
THIS CUSTOMER OBSERVED A DISCORDANT, NEGATIVE VITROS BENZ RESULT (<85 (NEGATIVE) VS. AN EXPECTED RESULT = 1015 NG/ML (POSITIVE)) FROM A SINGLE PATIENT SAMPLE USING THE VITROS 5,1 FS CHEMISTRY SYSTEM. THE AFFECTED RESULT WAS REPORTED OUT OF THE LABORATORY. BIASED PATIENT RESULTS OF THE DIRECTION AND MAGNITUDE OBSERVED COULD LEAD TO INAPPROPRIATE PHYSICIAN ACTION. THE PHYSICIAN QUESTIONED THE RESULT AS A POSITIVE TEST RESULT WAS EXPECTED. THE SAMPLE WAS SENT FOR CONFIRMATORY TESTING USING A GC/MS METHOD. A CORRECTED REPORT WAS ISSUED. THERE WAS NO REPORT OF PATIENT HARM AS A RESULT OF THIS EVENT. (B)(4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 226809 | VITROS CHEMISTRY PRODUCTS BENZ REAGENT | IN VITRO DIAGNOSTIC | JXM | ORTHO-CLINICAL DIAGNOSTICS | 1523-15-2524 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |