FDA Adverse Event
Malfunction
Summary report: N
9800
MDR report key: 3124056
·
Received May 17, 2013
Report
- Report Number
- 1720753-2013-06175
- Event Type
- Malfunction
- Date Received
- May 17, 2013
- Date of Event
- May 7, 2013
- Report Date
- May 17, 2013
- Manufacturer
- GE OEC MEDICAL SYSTEMS (SLC)
- Product Code
- JAA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A GE SVC REP PERFORMED AN ON SITE INVESTIGATION. THE POWER SUPPLY WAS REPLACED DURING THE SVC CALL. THE SYS WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK INTO SVC.
Description of Event or Problem · 1
IT WAS REPORTED THE SYS HAD AN INTERMITTENT COMMUNICATION FAILED ERROR MESSAGES. THIS ERROR RESULTS IN A SYS LOCK UP, NO BOOT, OR SHUT DOWN SITUATION. THERE IS NO REPORT OF INJURY OR DEATH ASSOCIATED WITH THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 219671 | 9800 | FLUOROSCOPIC X-RAY | JAA | GE OEC MEDICAL SYSTEMS (SLC) | 9800 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |