FDA Adverse Event Malfunction Summary report: N

HS III PROXIMAL SEAL SYSTEM 3.8MM

MDR report key: 3124016 · Received May 15, 2013

Report

Report Number
2242352-2013-00462
Event Type
Malfunction
Date Received
May 15, 2013
Date of Event
April 10, 2013
Report Date
April 25, 2013
Manufacturer
MAQUET CARDIOVASCULAR, LLC
Product Code
DXC
PMA / PMN Number
K080169
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TN, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE HAS NOT YET BEEN RETURNED TO THE FACTORY FOR EVAL. VISUAL INSPECTION DETERMINED NO SIGNS OF CLINICAL USAGE OR EVIDENCE OF BLOOD. THE DELIVERY DEVICE WAS NOT RETURNED. THE TENSION SPRING ASSEMBLY, THE ANCHOR TAB, AND THE SEAL WERE LOCATED INSIDE THE BODY OF THE LOADING DEVICE. BASED UPON THE RECEIVED CONDITION OF THE DEVICE, THE REPORTED COMPLAINT WAS CONFIRMED. A LOT HISTORY RECORD REVIEW WAS COMPLETED FOR THE REPORTED PRODUCT LOT NUMBER. THERE WAS NO NONCONFORMANCE RECORDED IN THE LOT HISTORY. (B)(4).

Description of Event or Problem · 1

THE HOSPITAL REPORTED THAT DURING A CORONARY ARTERY BYPASS PROCEDURE, THE HEARTSTRING III SEAL FAILED TO LOAD PROPERLY. A REPLACEMENT DEVICE WAS USED TO COMPLETE THE PROCEDURE. THE HOSPITAL DID NOT REPORT ANY PT EFFECTS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
214549 HS III PROXIMAL SEAL SYSTEM 3.8MM CLAMPLESS BEATING HEART DXC MAQUET CARDIOVASCULAR, LLC HSK-3038 25071704

Patients

Seq Age Sex Outcome Treatment
1 NA