HS III PROXIMAL SEAL SYSTEM 3.8MM
Report
- Report Number
- 2242352-2013-00463
- Event Type
- Malfunction
- Date Received
- May 15, 2013
- Date of Event
- April 11, 2013
- Report Date
- April 25, 2013
- Manufacturer
- MAQUET CARDIOVASCULAR, LLC
- Product Code
- DXC
- PMA / PMN Number
- K080169
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TN, US
- Reporter Occupation
- PHYSICIAN
Narratives
THE DEVICE HAS NOT YET BEEN RETURNED TO THE FACTORY FOR EVAL. IT SHOWED NO SIGNS OF CLINICAL USAGE OR EVIDENCE OF BLOOD. THE LOADING DEVICE WAS NOT RETURNED. THE TENSION SPRING ASSEMBLY AND ANCHOR TAB WERE LOCATED INSIDE THE TUBE OF THE DELIVERY DEVICE. THE SEAL WAS EXTENDED OUTSIDE OF THE DELIVERY TUBE; IT REMAINED ANCHORED TO THE TENSION SPRING ASSEMBLY. THE GREEN SLIDE LOCK WAS LOCKED AND WHITE PLUNGER WAS NOT DEPRESSED ON THE DELIVERY DEVICE. BASED UPON THE EVAL RESULTS, THE REPORTED COMPLAINT WAS CONFIRMED. A LOT HISTORY RECORD REVIEW WAS COMPLETED FOR THE REPORTED PRODUCT LOT NUMBER. THERE WAS NO NONCONFORMANCE RECORDED IN THE LOT HISTORY. (B)(4).
THE HOSPITAL REPORTED THAT DURING A CORONARY ARTERY HARVESTING PROCEDURE, THE HEARTSTRING III SEAL FAILED TO LOAD PROPERLY. A REPLACEMENT DEVICE WAS USED TO COMPLETE THE PROCEDURE. THE HOSPITAL DID NOT REPORT ANY PT EFFECTS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 214845 | HS III PROXIMAL SEAL SYSTEM 3.8MM | CLAMPLESS BEATING HEART | DXC | MAQUET CARDIOVASCULAR, LLC | HSK-3038 | 25064921 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA |