FDA Adverse Event
Malfunction
Summary report: N
INTRAMATIC LUX3 CONTRA ANGLE 25LHA
MDR report key: 3123997
·
Received May 15, 2013
Report
- Report Number
- 3003637274-2013-00016
- Event Type
- Malfunction
- Date Received
- May 15, 2013
- Date of Event
- February 15, 2013
- Report Date
- April 12, 2013
- Manufacturer
- KAVO DENTAL GMBH
- Product Code
- EFB
- PMA / PMN Number
- K073478
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WA, US
- Reporter Occupation
- DENTIST
Narratives
Additional Manufacturer Narrative · 1
THE ANALYSIS DID SHOW THAT THE BACK CAP BUTTON ON HANDPIECE COVER IS MISSING. FUNCTIONAL PROBLEMS OR DENTS ON THE PRODUCT WAS NOT FOUND. THE FINAL ROOT CAUSE CANNOT BE IDENTIFIED, BECAUSE THE BACK CAP BUTTON IS MISSING. FACTOR THAT MAY CONTRIBUTE TO THE BACK CAP BECOMING LOOSE ARE EXCESSIVE HANDLING OF THE PRODUCT, WEAR, ETC. THE PRODUCT WAS MORE THAN 10 YEAR IN USE. THE USER INSTRUCTION REQUIRES A VISUAL AND FUNCTIONAL TEST PRIOR TO EACH TREATMENT TO ENSURE THAT THE HANDPIECE IS RUNNING WITHIN SPEC.
Description of Event or Problem · 1
DURING A STANDARD DENTAL TREATMENT THE BACK CAP OF A HANDPIECE CAME OFF AND WAS SWALLOWED BY A PT. NO MEDICAL CARE WAS GIVEN TO PT. (B)(6) (DENTIST ASSISTANT) SAID PT WAS FINE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 214545 | INTRAMATIC LUX3 CONTRA ANGLE 25LHA | DENTAL HANDPIECE | EFB | KAVO DENTAL GMBH | 25LHA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 40 YR |