FDA Adverse Event Malfunction Summary report: N

GENTLE POWER LUX CONTRA ANGLE 25LPA

MDR report key: 3123996 · Received May 15, 2013

Report

Report Number
3003637274-2013-00014
Event Type
Malfunction
Date Received
May 15, 2013
Date of Event
April 8, 2013
Report Date
April 8, 2013
Manufacturer
KAVO DENTAL GMBH
Product Code
EFB
PMA / PMN Number
K073478
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
DENTIST

Narratives

Additional Manufacturer Narrative · 1

THE ANALYSIS DID SHOW VERY HEAVY RESIDUE IN THE HANDPIECE, THE BEARING WAS STIFF. THE STIFF BEARING CAUSED HIGH FRICTION AND THEREFORE INCREASE OF TEMPERATURE. THE CONDITION OF THE HANDPIECE INDICATES THAT THE MAINTENANCE PROCESSING IS NOT CARRIED OUT AS REQUESTED. THE USER INSTRUCTION REQUIRES A VISUAL AND FUNCTIONAL TEST PRIOR TO EACH TREATMENT TO ENSURE THAT THE HANDPIECE IS RUNNING WITHIN SPEC. IT CONTAINS ALSO A NOTE THAT THE HANDPIECE SHOULD NOT TOUCH SOFT TISSUE DURING THE TREATMENT. REPORTED DUE TO THE 2 YEAR PRESUMPTION RULE.

Description of Event or Problem · 1

DURING A STANDARD DENTAL TREATMENT, WHILE WORKING ON TOOTH #30 THE PT WAS BURNED ON THE TONGUE. THE BURN IS ABOUT 4-5 MM. NOTHING WAS PRESCRIBED OR GIVEN TO PT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
214978 GENTLE POWER LUX CONTRA ANGLE 25LPA DENTAL HANDPIECE EFB KAVO DENTAL GMBH 25LPA

Patients

Seq Age Sex Outcome Treatment
1 50 YR