FDA Adverse Event
Malfunction
Summary report: N
GENTLE POWER LUX CONTRA ANGLE 25LPA
MDR report key: 3123996
·
Received May 15, 2013
Report
- Report Number
- 3003637274-2013-00014
- Event Type
- Malfunction
- Date Received
- May 15, 2013
- Date of Event
- April 8, 2013
- Report Date
- April 8, 2013
- Manufacturer
- KAVO DENTAL GMBH
- Product Code
- EFB
- PMA / PMN Number
- K073478
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- DENTIST
Narratives
Additional Manufacturer Narrative · 1
THE ANALYSIS DID SHOW VERY HEAVY RESIDUE IN THE HANDPIECE, THE BEARING WAS STIFF. THE STIFF BEARING CAUSED HIGH FRICTION AND THEREFORE INCREASE OF TEMPERATURE. THE CONDITION OF THE HANDPIECE INDICATES THAT THE MAINTENANCE PROCESSING IS NOT CARRIED OUT AS REQUESTED. THE USER INSTRUCTION REQUIRES A VISUAL AND FUNCTIONAL TEST PRIOR TO EACH TREATMENT TO ENSURE THAT THE HANDPIECE IS RUNNING WITHIN SPEC. IT CONTAINS ALSO A NOTE THAT THE HANDPIECE SHOULD NOT TOUCH SOFT TISSUE DURING THE TREATMENT. REPORTED DUE TO THE 2 YEAR PRESUMPTION RULE.
Description of Event or Problem · 1
DURING A STANDARD DENTAL TREATMENT, WHILE WORKING ON TOOTH #30 THE PT WAS BURNED ON THE TONGUE. THE BURN IS ABOUT 4-5 MM. NOTHING WAS PRESCRIBED OR GIVEN TO PT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 214978 | GENTLE POWER LUX CONTRA ANGLE 25LPA | DENTAL HANDPIECE | EFB | KAVO DENTAL GMBH | 25LPA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 50 YR |