FDA Adverse Event
Malfunction
Summary report: N
11.0 CM LONG ATTACHMENT
MDR report key: 3123994
·
Received May 22, 2013
Report
- Report Number
- 1045834-2013-02165
- Event Type
- Malfunction
- Date Received
- May 22, 2013
- Date of Event
- April 18, 2013
- Report Date
- April 22, 2013
- Manufacturer
- DEPUY SYNTHES POWER TOOLS
- Product Code
- HBC
- PMA / PMN Number
- K011444
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OR, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
THE ATTACHMENT DEVICE WAS RECEIVED FOR EVALUATION. THE ATTACHMENT DEVICE WAS TESTED AND THE DEVICE HAS WORN/DAMAGED BEARINGS. EVIDENCE INDICATED THIS WAS DUE TO USAGE WEAR OVER TIME. THE REPORTED CONDITION WAS CONFIRMED. IF ADDITIONAL INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SENT. (B)(4).
Description of Event or Problem · 1
REPORT RECEIVED FROM THE USA REGARDING AN ATTACHMENT DEVICE IN WHICH, DURING INSPECTION/TESTING, IT WAS OBSERVED THAT THE DEVICE WAS MAKING A "LOUD NOISE". THE ATTACHMENT DEVICE WAS NOT USED IN SURGERY. NO INJURIES OR MEDICAL INTERVENTION WERE REPORTED. NO ADDITIONAL INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 225782 | 11.0 CM LONG ATTACHMENT | MOTOR, DRILL, ELECTRIC- ATTACHMENT | HBC | DEPUY SYNTHES POWER TOOLS |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |