FDA Adverse Event Malfunction Summary report: N

GENTLE POWER LUX CONTRA ANGLE 25LPR

MDR report key: 3123939 · Received May 15, 2013

Report

Report Number
3003637274-2013-00017
Event Type
Malfunction
Date Received
May 15, 2013
Date of Event
April 8, 2013
Report Date
April 18, 2013
Manufacturer
KAVO DENTAL GMBH
Product Code
EFB
PMA / PMN Number
K073478
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AZ, US
Reporter Occupation
DENTIST

Narratives

Additional Manufacturer Narrative · 1

THE ANALYSIS DID SHOW THAT THE HANDPIECE RUNS OUT OF SPEC. FURTHER, THE BACK CAP BUTTON STICKS IN DUE TO DENTED HEAD. THIS CONDITION AFFECTED THE FUNCTIONALITY OF THE DRIVE AND THE BACK UP BUTTON. CAUSED BY INNER FRICTION THE HANDPIECE HEAT UP. THE CONDITION OF THE HANDPIECE INDICATES THAT THE MAINTENANCE PROCESSING IS NOT CARRIED OUT AS REQUESTED. THE USER INSTRUCTION REQUIRES A VISUAL AND FUNCTIONAL TEST PRIOR TO EACH TREATMENT TO ENSURE THAT THE HANDPIECE IS RUNNING WITHIN SPEC. IT CONTAINS ALSO A NOTE THAT THE HANDPIECE SHOULD NOT TOUCH SOFT TISSUE DURING THE TREATMENT. REPORTED DUE TO THE 2 YEAR PRESUMPTION RULE.

Description of Event or Problem · 1

DURING A STANDARD DENTAL TREATMENT ON TOOTH# 4, PT WAS BURNED AT CORNER LIP. THE BURN WAS APPROX THE SIZE OF THE BACK CAP OF AN HANDPIECE. NO OINTMENT OR MEDICATIONS MENTIONED IN CHART.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
215325 GENTLE POWER LUX CONTRA ANGLE 25LPR DENTAL HANDPIECE EFB KAVO DENTAL GMBH 25LPR

Patients

Seq Age Sex Outcome Treatment
1 49 YR