FDA Adverse Event Malfunction Summary report: N

9900

MDR report key: 3123937 · Received May 17, 2013

Report

Report Number
1720753-2013-06145
Event Type
Malfunction
Date Received
May 17, 2013
Date of Event
May 7, 2013
Report Date
May 17, 2013
Manufacturer
GE OEC MEDICAL SYSTEMS
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REP PERFORMED AN ONSITE INVESTIGATION. THE HIGH VOLTAGE CABLE WAS REPLACED. THE SYSTEM WAS THEN FOUND TO BE OPERATING AS INTENDED.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE SYSTEM DISPLAYED X-RAY DISABLED AND COMMUNICATION ERROR MESSAGES. THESE ERROR MESSAGES LIKELY RESULTED IN A LOSS OF X-RAY FUNCTIONALITY. THERE IS NO REPORT OF PATIENT INJURY ASSOCIATED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
218921 9900 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS 9900

Patients

Seq Age Sex Outcome Treatment
1