FDA Adverse Event Malfunction Summary report: N

9900

MDR report key: 3123932 · Received May 17, 2013

Report

Report Number
1720753-2013-06135
Event Type
Malfunction
Date Received
May 17, 2013
Date of Event
May 10, 2013
Report Date
May 17, 2013
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REPRESENTATIVE PERFORMED AN ONSITE INVESTIGATION. THE INTERCONNECT CABLE CONNECTOR WAS RETIGHTENED. THE SYSTEM WAS THEN FOUND TO BE OPERATING AS INTENDED.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE IMAGE WENT WHITE, RESULTING IN A LOSS OF THE LIVE IMAGE. THERE IS NO REPORT OF PT INJURY ASSOCIATED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
219431 9900 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9900

Patients

Seq Age Sex Outcome Treatment