FDA Adverse Event
Malfunction
Summary report: N
XMAX MOTOR, HIGH STRENGTH HOSE
MDR report key: 3123930
·
Received May 22, 2013
Report
- Report Number
- 1045834-2013-02161
- Event Type
- Malfunction
- Date Received
- May 22, 2013
- Date of Event
- March 26, 2013
- Report Date
- April 22, 2013
- Manufacturer
- DEPUY SYNTHES POWER TOOLS
- Product Code
- ERL
- PMA / PMN Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE MOTOR DEVICE WAS RECEIVED FOR EVALUATION. THE MOTOR DEVICE WAS TESTED AND IT A LEAK WAS OBSERVED NEAR THE PRESSURE RELEASE VALVE. EVIDENCE INDICATES THIS IS DUE TO USAGE WEAR OVER TIME. THE REPORTED CONDITION WAS CONFIRMED. IF ADDITIONAL INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SENT. (B)(4).
Description of Event or Problem · 1
REPORT RECEIVED FROM USA REGARDING A MOTOR DEVICE IN WHICH A HOLE WAS OBSERVED IN THE "HOSE NEAR THE HANDPIECE" AND "IS LEAKING OIL". THE MOTOR DEVICE WAS NOT USED IN SURGERY. NO INJURIES OR MEDICAL INTERVENTION WERE REPORTED. NO ADDITIONAL INFORMATION IS AVAIABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 226095 | XMAX MOTOR, HIGH STRENGTH HOSE | DRILL, SURGICAL, ENT (PNEUMATIC)- HANDPIECE | ERL | DEPUY SYNTHES POWER TOOLS |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |