FDA Adverse Event Malfunction Summary report: N

EMAX 2 PLUS MOTOR

MDR report key: 3123918 · Received May 22, 2013

Report

Report Number
1045834-2013-02159
Event Type
Malfunction
Date Received
May 22, 2013
Date of Event
April 22, 2013
Report Date
April 22, 2013
Manufacturer
DEPUY SYNTHES POWER TOOLS
Product Code
HBC
PMA / PMN Number
K080802
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE MOTOR DEVICE WAS RECEIVED FOR EVALUATION. THE MOTOR DEVICE WAS TESTED AND MET TEMPERATURE SPECIFICATIONS. THE REPORTED CONDITION WAS NOT CONFIRMED. (B)(4).

Description of Event or Problem · 1

REPORT RECEIVED FROM THE USA REGARDING A MOTOR DEVICE IN WHICH, DURING ROUTINE INSPECTION, IT AS DISCOVERED THAT THE MOTOR DEVICE WAS "OVERHEATING" AND "GETTING WARM TO THE TOUCH". THE MOTOR DEVICE WAS NOT USED IN SURGERY. NO INJURIES OR MEDICAL INTERVENTION WERE REPORTED. NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
226077 EMAX 2 PLUS MOTOR MOTOR, DRILL, ELECTRIC- HANDPIECE HBC DEPUY SYNTHES POWER TOOLS

Patients

Seq Age Sex Outcome Treatment
1