LIBERTY CYCLER SET, SINGLE CONN./ EXT. DL
Report
- Report Number
- 8030665-2013-00286
- Event Type
- Malfunction
- Date Received
- May 15, 2013
- Date of Event
- April 16, 2013
- Report Date
- April 16, 2013
- Manufacturer
- REYNOSA MANUFACTURING
- Product Code
- FKX
- PMA / PMN Number
- K043363
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- PATIENT
Narratives
THE DEVICE WAS NOT RETURNED TO THE MANUFACTURER FOR PHYSICAL EVALUATION AND THE FAILURE MODE CANNOT BE CONFIRMED. HOWEVER, AN INVESTIGATION OF THE DEVICE MANUFACTURING RECORDS WAS CONDUCTED BY THE MANUFACTURER. THERE WERE NO DEVIATIONS OR NONCONFORMANCES DURING THE MANUFACTURING PROCESS. IN ADDITION, THE BATCH RECORD REVIEW CONFIRMED THE LABELING, MATERIAL, AND PROCESS CONTROLS WERE WITHIN SPECIFICATION.
A PERITONEAL DIALYSIS PATIENT HAS REPORTED THAT DIALYSIS SOLUTION WAS LEAKING OUT OF THE CASSETTE AND INTO THE CYCLER. PATIENT WAS IN DRAIN THREE OF TREATMENT. UPON REMOVING THE TUBING SET, FLUID WAS FOUND LEAKING INTO THE CYCLER. PATIENT DID NOT RECEIVE ANY ANTIBIOTICS AND HAS HAD NO ADVERSE EFFECTS. SAMPLE WAS DISCARDED BY THE PATIENT, SAMPLE IS NOT AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 215021 | LIBERTY CYCLER SET, SINGLE CONN./ EXT. DL | FKX | REYNOSA MANUFACTURING | 13BR08832 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | LIBERTY DIALYSIS CYCLER |