FDA Adverse Event Malfunction Summary report: N

840 VENTILATOR

MDR report key: 3123847 · Received May 15, 2013

Report

Report Number
8020893-2013-01085
Event Type
Malfunction
Date Received
May 15, 2013
Date of Event
March 25, 2013
Report Date
April 17, 2013
Manufacturer
COVIDIEN, FORMERLY NELLCOR PURITAN
Product Code
CBK
PMA / PMN Number
K970460
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

(B)(4).

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT AN 840 VENTILATOR WAS INOPERABLE. THERE WAS NO PT INVOLVEMENT. THE COVIDIEN CUSTOMER SUPPORT ENGINEER (CSE) VERIFIED THE MALFUNCTION. THE CSE INSPECTED THE DEVICE AND REPLACED THE ANALOG INTERFACE (AI) PCB. THE UNIT PASSED EXTENDED SELF-TESTING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
214987 840 VENTILATOR VENTILATOR CBK COVIDIEN, FORMERLY NELLCOR PURITAN 840

Patients

Seq Age Sex Outcome Treatment
1