KNIFE
Report
- Report Number
- 2523835-2013-00065
- Event Type
- Malfunction
- Date Received
- May 15, 2013
- Report Date
- April 18, 2013
- Manufacturer
- ALCON PRECISION DEVICE - SINKING SPRING
- Product Code
- HNN
- PMA / PMN Number
- NA
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SC, US
- Reporter Occupation
- OTHER
Narratives
ONE OPENED SAMPLE WAS RETURNED. THE SAMPLE WAS EXAMINED USING 10X MAGNIFICATION AND FOUND TO HAVE A DAMAGED TIP AND CUTTING EDGES. THE DEVICE HISTORY RECORD (DHR) FOR THE LOT WAS REVIEWED. FUNCTIONAL PENETRATION TEST VALUES FOR THIS LOT MET SPECIFICATION. NO ABNORMALITIES THAT COULD HAVE CONTRIBUTED TO A DULL KNIFE WERE FOUND DURING THE DHR REVIEW AND THE PRODUCT WAS RELEASED ACCORDING TO THE MANUFACTURER'S ACCEPTANCE CRITERIA. THE EXACT ROOT CAUSE FOR THE DAMAGED KNIVES IS UNKNOWN; HOW OR WHEN THE BLADES BECAME DAMAGED CANNOT BE DETERMINED. THE DAMAGE TO THE RETURNED SAMPLE IS CONSISTENT WITH DAMAGE THAT CAN OCCUR WHEN THE BLADE CONTACTS ANOTHER SURFACE PRIOR TO USE. SOME POTENTIAL CAUSES COULD BE REUSE, IMPROPER HANDLING, OR CONTACT WITH ANOTHER INSTRUMENT ON THE INSTRUMENT TRAY DURING PROCEDURE SETUP. ALL KNIVES ARE 100% INSPECTED BY TRAINED OPERATORS USING A MINIMUM OF 10X MAGNIFICATION DURING MANUFACTURING. ANY DEFECTS, SUCH AS THE DAMAGED TIP AND CUTTING EDGES EXHIBITED ON THE RETURNED OPENED SAMPLE, ARE REMOVED FROM THE LOT AND SCRAPPED. PENETRATION TESTING IS PERFORMED AND MONITORED DURING THE FINISHING PROCESS TO ENSURE THE SHARPNESS OF THE PRODUCT. (B)(4).
A NURSE REPORTED THAT THEY HAVE HAD BLADES THAT WERE DULL OR BENT AT THE TIP. EACH TIME THE ISSUE HAS BEEN NOTED, THE BLADES WERE IMMEDIATELY REPLACED WITH NO PROCEDURAL DELAY. THERE HAS BEEN NO PATIENT IMPACT. THIS IS THE FIRST OF THREE REPORTS FOR THIS FACILITY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 214494 | KNIFE | MANUAL OPHTHALMIC SURGICAL INSTRUMENT | HNN | ALCON PRECISION DEVICE - SINKING SPRING | 8065921540 | 917046M |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |