SCREWDRIVER AXSOS T15 4.0MM LOCKING SET
Report
- Report Number
- 0008031020-2013-00159
- Event Type
- Malfunction
- Date Received
- May 22, 2013
- Date of Event
- February 23, 2013
- Report Date
- April 25, 2013
- Manufacturer
- STRYKER TRAUMA SELZACH
- Product Code
- LXH
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OK, US
- Reporter Occupation
- OTHER
Narratives
ONCE THE INVESTIGATION HAS BEEN COMPLETED ANY ADDITIONAL INFORMATION WILL BE REPORTED IN A SUPPLEMENTAL REPORT.
EVALUATION SUMMARY: THE REPORTED EVENT THAT THE SCREWDRIVER TIP BROKE COULD BE CONFIRMED. BASED ON THE INVESTIGATION, THE ROOT CAUSE OF THE DETERMINED BROKEN TIP WAS ATTRIBUTED TO A USER RELATED ISSUE. THE BREAKAGE WAS CAUSED BY OVER TORQUE. A REVIEW OF THE LABELING DID NOT INDICATE ANY ABNORMALITIES. A REVIEW OF THE DEVICE HISTORY FOR THE REPORTED LOT DID NOT INDICATE ANY ABNORMALITIES. NO CORRECTIVE ACTIONS ARE REQUIRED AT THIS TIME. INDICATIONS FOR ANY MATERIAL, MANUFACTURING OR DESIGN RELATED PROBLEMS WERE NOT DETERMINED IN THE INVESTIGATION. PLEASE NOTE THAT THE CURRENT OPTECH STATES: FINAL TIGHTENING OF LOCKING SCREWS SHOULD ALWAYS BE PERFORMED MANUALLY USING THE TORQUE LIMITING ATTACHMENT.
TIP OF THE SCREWDRIVER BROKE WHEN INSERTING A AXSOS 4.0 LOCKING SCREW.
TIP OF THE SCREWDRIVER BROKE WHEN INSERTING A AXSOS 4.0 LOCKING SCREW.