EXTRACTION ROD, CONICAL T2 HUMERUS
Report
- Report Number
- 0009610622-2013-00276
- Event Type
- Malfunction
- Date Received
- May 22, 2013
- Date of Event
- April 29, 2013
- Report Date
- April 30, 2013
- Manufacturer
- STRYKER OSTEOSYNTHESIS-KIEL
- Product Code
- LXH
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AU
- Reporter Occupation
- NURSE
Narratives
ONCE THE INVESTIGATION HAS BEEN COMPLETED, ANY ADDITIONAL INFORMATION WILL BE REPORTED IN A SUPPLEMENTAL REPORT.
EVALUATION SUMMARY: THE REPORTED ISSUE WAS CONFIRMED. EVALUATION REVEALED BOTH INSTRUMENTS TO BE PRIMARY PRODUCTS. NO ASSOCIATED PRODUCTS WERE REPORTED. NO DEVIATIONS WERE FOUND DURING REVIEW OF THE MANUFACTURING AND INSPECTION DOCUMENTS (DHR). THE ITEMS RETURNED WERE DOCUMENTED AS FAULTLESS PRIOR TO DISTRIBUTION. AS THE DEVICES HAD BEEN IN USE FOR YEARS WE PRE-SUPPOSE THAT THEY HAD FULFILLED THEIR TASKS IN FORMER SURGERIES AS INTENDED. THE DAMAGED THREADS INDICATE THAT BOTH INSTRUMENTS WERE INSERTED IN OBLIQUE MODE AND CROSS THREADING OCCURRED. BOTH INSTRUMENTS ARE MADE OF THE SAME MATERIAL AND WERE BOTH HARDENED TO APPROX. 47 HRC, THEREFORE IT COULD NOT BE DETERMINED WHICH INSTRUMENT CAUSED THE THREAD DAMAGES. IT IS POSSIBLE THAT BOTH THREADS WERE FULLY FUNCTIONAL PRIOR THE ASSEMBLING BUT A PRE-DAMAGING COULD ALSO NOT BE EXCLUDED. HOWEVER, INSERTION IN AN OBLIQUE MODE IS CLASSIFIED AS USER ERROR. DURING THE INVESTIGATION OF THE EXTRACTION ROD, CONICAL THE LASER-MARKING ¿MUSTER¿ WAS FOUND ON THE THICKER SHAFT PART. MUSTER MEANS SAMPLE. THEREFORE THE USER USED A SAMPLE INSTRUMENT. SAMPLE INSTRUMENTS ARE MADE ONLY FOR VISUALIZATIONS LIKE WORKSHOPS OR TRAININGS; THEREFORE IT IS NOT ALLOWED TO USE THEM DURING SURGERIES. HOW THE SAMPLE EXTRACTION ROD, CONICAL GOT TO THE HOSPITAL COULD NOT BE CLARIFIED.
IT WAS REPORTED THAT THE THREADS HAVE BECOME WEDGED INTO EACH OTHER.
IT WAS REPORTED THAT THE THREADS HAVE BECOME WEDGED INTO EACH OTHER.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 225586 | EXTRACTION ROD, CONICAL T2 HUMERUS | INSTRUMENT | LXH | STRYKER OSTEOSYNTHESIS-KIEL | K639978 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |