EON MINI
Report
- Report Number
- 1627487-2013-12695
- Event Type
- Injury
- Date Received
- May 17, 2013
- Date of Event
- April 24, 2013
- Report Date
- April 24, 2013
- Manufacturer
- ST. JUDE MEDICAL - NEUROMODULATION
- Product Code
- LGW
- PMA / PMN Number
- P010032
- Removal / Correction Number
- 1627487-07262012-002-R 1
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
THIS IPG SERIAL NUMBER WAS INCLUDED IN A FIELD ADVISORY. THIS MODEL NUMBER WAS INCLUDED IN A FIELD CORRECTION. SJM HAS LIMITED INFO RELATED TO THE PT'S MEDICAL HISTORY AND IS UNABLE TO FORM AN OPINION AS TO THE RELEVANCY OF THE PT'S HISTORY TO THE EVENT REPORTED. SJM DEFERS TO THE PT'S PHYSICIAN REGARDING MEDICAL HISTORY.
DEVICE 1 OF 2. REFERENCE MFR REPORT #1627487-2013-12696. IT WAS REPORTED THE PT EXPERIENCED POCKET HEATING. THE PT REPORTED WITHIN 10 MINS OF CHARGING, THE SITE WOULD BECOME READ AND UNCOMFORTABLY HOT AND HER SKIN WAS RED. DUE TO THE ISSUE, THE PT STOPPED CHARGING THE IPG. AS A RESULT, THE BATTERY IS DEPLETED. THE PT PLANS TO UNDERGO SURGICAL INTERVENTION TO ADDRESS THE ISSUE. ON (B)(4) 2012 ST. JUDE MEDICAL, NEUROMODULATION DIV, SENT FIELD ACTION LETTERS TO PTS RELATED TO HEATING WHILE CHARGING AND RAISED AWARENESS OF THIS ISSUE TO PTS. AN INCREASE IN PRIOR NON-REPORTED HEATING WHILE CHARGING EVENTS AND OTHER NON-REPORTED EVENTS WAS EXPECTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 218950 | EON MINI | SCS IPG | LGW | ST. JUDE MEDICAL - NEUROMODULATION | 3788 | 3315469 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 72 YR | Required Intervention | IMPLANT:| SCS LEAD: MODEL 3245 |