MARYLAND BIPOLAR FORCEPS INSTRUMENT
Report
- Report Number
- 2955842-2013-01770
- Event Type
- Malfunction
- Date Received
- May 22, 2013
- Date of Event
- April 18, 2013
- Report Date
- April 23, 2013
- Manufacturer
- INTUITIVE SURGICAL,INC.
- Product Code
- NAY
- PMA / PMN Number
- K050369
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- KS
- Reporter Occupation
- OTHER
Narratives
THE INSTRUMENT WAS RETURNED AND EVALUATED. ENGINEERING CONFIRMED THE REPORTED COMPLAINT. FOR CLARIFICATION, THE INSTRUMENT WAS FOUND WITH A FRAYED PITCH CABLE AT DISTAL CLEVIS. NO DAMAGE WAS FOUND ON THE PULLEY AND CLEVIS. ADDITIONAL DAMAGE FOUND WAS A CORRODED BACK IDLER PULLEY. ON THE SURFACE OF EACH PULLEY EXHIBITED A YELLOWISH MATERIAL. ENGINEERING CONCLUDED THE DAMAGE WAS LIKELY DUE TO IMPROPER CLEANING. ELECTRICAL CONTINUITY PASSED. NO OTHER DAMAGE WAS FOUND. THE INSTRUMENTS AND ACCESSORIES USER MANUAL SPECIFICALLY STATES: WARNING: READ ALL INSTRUCTIONS CAREFULLY. FAILURE TO PROPERLY FOLLOW INSTRUCTIONS MAY CAUSE IMPROPER FUNCTIONING OF THE DEVICE. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A MDR REPORTABLE EVENT; HOWEVER; THE REPORTED MALFUNCTION IF TO RECUR COULD CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT.
IT WAS REPORTED THAT AFTER A DA VINCI SI TOTAL BENIGN HYSTERECTOMY PROCEDURE A BROKEN CABLE WAS IDENTIFIED ON A MARYLAND BIPOLAR FORCEPS INSTRUMENT. THERE WERE NO MISSING PIECES OR FALLEN PIECES REPORTED. THE PLANNED SURGICAL PROCEDURE WAS COMPLETED AND NO PATIENT HARM, ADVERSE OUTCOME, OR INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 225976 | MARYLAND BIPOLAR FORCEPS INSTRUMENT | ENDOSCOPIC ELECTROSURGICAL INSTRUMENT | NAY | INTUITIVE SURGICAL,INC. | 420172-07 | M10121215 886 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | DA VINCI SI SYSTEM INSTRUMENTS & ACCESSORIES ESU |