FDA Adverse Event Malfunction Summary report: N

MARYLAND BIPOLAR FORCEPS INSTRUMENT

MDR report key: 3123474 · Received May 22, 2013

Report

Report Number
2955842-2013-01770
Event Type
Malfunction
Date Received
May 22, 2013
Date of Event
April 18, 2013
Report Date
April 23, 2013
Manufacturer
INTUITIVE SURGICAL,INC.
Product Code
NAY
PMA / PMN Number
K050369
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KS
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSTRUMENT WAS RETURNED AND EVALUATED. ENGINEERING CONFIRMED THE REPORTED COMPLAINT. FOR CLARIFICATION, THE INSTRUMENT WAS FOUND WITH A FRAYED PITCH CABLE AT DISTAL CLEVIS. NO DAMAGE WAS FOUND ON THE PULLEY AND CLEVIS. ADDITIONAL DAMAGE FOUND WAS A CORRODED BACK IDLER PULLEY. ON THE SURFACE OF EACH PULLEY EXHIBITED A YELLOWISH MATERIAL. ENGINEERING CONCLUDED THE DAMAGE WAS LIKELY DUE TO IMPROPER CLEANING. ELECTRICAL CONTINUITY PASSED. NO OTHER DAMAGE WAS FOUND. THE INSTRUMENTS AND ACCESSORIES USER MANUAL SPECIFICALLY STATES: WARNING: READ ALL INSTRUCTIONS CAREFULLY. FAILURE TO PROPERLY FOLLOW INSTRUCTIONS MAY CAUSE IMPROPER FUNCTIONING OF THE DEVICE. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A MDR REPORTABLE EVENT; HOWEVER; THE REPORTED MALFUNCTION IF TO RECUR COULD CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT AFTER A DA VINCI SI TOTAL BENIGN HYSTERECTOMY PROCEDURE A BROKEN CABLE WAS IDENTIFIED ON A MARYLAND BIPOLAR FORCEPS INSTRUMENT. THERE WERE NO MISSING PIECES OR FALLEN PIECES REPORTED. THE PLANNED SURGICAL PROCEDURE WAS COMPLETED AND NO PATIENT HARM, ADVERSE OUTCOME, OR INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
225976 MARYLAND BIPOLAR FORCEPS INSTRUMENT ENDOSCOPIC ELECTROSURGICAL INSTRUMENT NAY INTUITIVE SURGICAL,INC. 420172-07 M10121215 886

Patients

Seq Age Sex Outcome Treatment
1 DA VINCI SI SYSTEM INSTRUMENTS & ACCESSORIES ESU