OT VERIO IQ METER
Report
- Report Number
- 3008382007-2013-12474
- Event Type
- Malfunction
- Date Received
- May 21, 2013
- Report Date
- April 25, 2013
- Manufacturer
- LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL
- Product Code
- NBW
- PMA / PMN Number
- K110637
- Removal / Correction Number
- 2013-IE-017
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- OTHER
Narratives
FOLLOW-UP # 2 ¿ (07/15/2013). THE PATIENT¿S METER HAS BEEN RETURNED AND EVALUATED BY LIFESCAN PRODUCT ANALYSIS ON 6/21/2013 AND 7/11/2013, RESPECTIVELY, WITH THE FOLLOWING FINDINGS:THE METER PASSED ALL TESTING WITH NO FAULTS FOUND. THE REPORTED ISSUE COULD NOT BE REPRODUCED. IF LIFESCAN OBTAINS ADDITIONAL INFORMATION REGARDING THIS COMPLAINT, A FOLLOW UP REPORT WILL BE SUBMITTED. AT THIS TIME, LIFESCAN CONSIDERS THIS MATTER CLOSED.
FOLLOW-UP # 1 ¿ (07/05/2013). THE PATIENT¿S TEST STRIPS HAVE BEEN RETURNED AND EVALUATED BY LIFESCAN PRODUCT ANALYSIS ON 6/12/2013 AND 6/28/2013, RESPECTIVELY, WITH THE FOLLOWING FINDINGS:THE TEST STRIPS PASSED ALL TESTING WITH NO FAULTS FOUND. THE REPORTED ISSUE COULD NOT BE CONFIRMED. THE METER HAS ALSO BEEN RETURNED TO LFS; HOWEVER, THE INVESTIGATION HAS NOT BEEN COMPLETED. NO CONCLUSIONS CAN BE DRAWN AT THIS TIME. ONCE THE EVALUATION HAS BEEN COMPLETED A SUPPLEMENTAL REPORT WILL BE FILED.IF LIFESCAN OBTAINS ADDITIONAL INFORMATION REGARDING THIS COMPLAINT, A FOLLOW UP REPORT WILL BE SUBMITTED. AT THIS TIME, LIFESCAN CONSIDERS THIS MATTER CLOSED.
LIFESCAN HAS REQUESTED THE SUBJECT PRODUCTS TO BE RETURNED FOR EVALUATION HOWEVER THE PRODUCTS HAVE NOT YET BEEN RETURNED OR EVALUATED. IF LIFESCAN RECEIVES THE PRODUCT LIFESCAN WILL INFORM THE FDA OF ANY PRODUCTS THAT FAIL EVALUATION IN A SUPPLEMENTAL REPORT.
ON (B)(6) 2013, THE REPORTER CONTACTED LIFESCAN (B)(4), ALLEGING INACCURATELY HIGH RESULTS. THE REPORTER MENTIONED READINGS OF 12.2 MMOL/L ON THE LIFESCAN METER AND 8.4 MMOL/L ON THE OTHER METER. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED RESULTS DID NOT MEET LIFESCAN¿S ACCURACY/PRECISION CRITERIA AND THE ALLEGED INACCURACY ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 223785 | OT VERIO IQ METER | GLUCOSE MONITORING SYS/KIT | NBW | LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL | 3376906 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 68 YR |