FDA Adverse Event
Malfunction
Summary report: N
PARADIGM REAL-TIME INSULIN INFUSION PUMP
MDR report key: 3123022
·
Received May 21, 2013
Report
- Report Number
- 3004209178-2013-93823
- Event Type
- Malfunction
- Date Received
- May 21, 2013
- Date of Event
- May 1, 2013
- Report Date
- May 1, 2013
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL
- Reporter Occupation
- PATIENT
Narratives
Additional Manufacturer Narrative · 1
THE INSULIN PUMP PASSED THE DISPLACEMENT TEST. THE INSULIN PUMP WAS UNABLE TO PRIME DURING PRIME TEST AND ALARMED MOTOR ERROR DURING BASIC OCCLUSION TEST DUE TO PROTRUDED/LOOSE DRIVE SUPPORT DISK. NO PRIME ALARM NOTED. MOTOR PASSED MOTOR TEST. THE INSULIN PUMP HAD A SCRATCHED DISPLAY WINDOW AND A BROKEN BELT CLIP SLOT.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE INSULIN PUMP RECEIVED MOTOR ERROR AND PRIME ALARMS. CUSTOMER'S BLOOD GLUCOSE READING IS 137 MG/DL. CUSTOMER STATES THAT DRIVE SUPPORT CAP IS PROTRUDED. THE DISPLACEMENT TEST FAILED. ADVISED CUSTOMER THAT THE INSULIN PUMP MUST BE REPLACED. NOTHING FURTHER REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 224606 | PARADIGM REAL-TIME INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-722NAP |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 64 YR |