FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 3123022 · Received May 21, 2013

Report

Report Number
3004209178-2013-93823
Event Type
Malfunction
Date Received
May 21, 2013
Date of Event
May 1, 2013
Report Date
May 1, 2013
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE INSULIN PUMP PASSED THE DISPLACEMENT TEST. THE INSULIN PUMP WAS UNABLE TO PRIME DURING PRIME TEST AND ALARMED MOTOR ERROR DURING BASIC OCCLUSION TEST DUE TO PROTRUDED/LOOSE DRIVE SUPPORT DISK. NO PRIME ALARM NOTED. MOTOR PASSED MOTOR TEST. THE INSULIN PUMP HAD A SCRATCHED DISPLAY WINDOW AND A BROKEN BELT CLIP SLOT.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE INSULIN PUMP RECEIVED MOTOR ERROR AND PRIME ALARMS. CUSTOMER'S BLOOD GLUCOSE READING IS 137 MG/DL. CUSTOMER STATES THAT DRIVE SUPPORT CAP IS PROTRUDED. THE DISPLACEMENT TEST FAILED. ADVISED CUSTOMER THAT THE INSULIN PUMP MUST BE REPLACED. NOTHING FURTHER REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
224606 PARADIGM REAL-TIME INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-722NAP

Patients

Seq Age Sex Outcome Treatment
1 64 YR