FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 3123018 · Received May 21, 2013

Report

Report Number
3004209178-2013-93814
Event Type
Malfunction
Date Received
May 21, 2013
Date of Event
May 1, 2013
Report Date
May 1, 2013
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

MOTOR ERROR ALARMED DURING REWIND AND ERROR ALARM FOUND IN HISTORY FILE DUE TO CORROSION ON THE MOTOR HOME SWITCH. NO MOISTURE DAMAGE FOUND ON ELECTRONIC ASSEMBLY. ALSO, THE INSULIN PUMP WAS RECEIVED WITH PROTRUDED/LOOSE DRIVE SUPPORT DISK.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE INSULIN PUMP ALARMED MOTOR ERROR. CUSTOMER'S BLOOD GLUCOSE READING IS 46 MG/DL. CUSTOMER TREATED WITH MANUAL INJECTION. THE DRIVE SUPPORT CAP IS PROTRUDED. ADVISED CUSTOMER THE INSULIN PUMP REQUIRES REPLACEMENT. NOTHING FURTHER REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
224578 PARADIGM REAL-TIME INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-722NAS

Patients

Seq Age Sex Outcome Treatment
1 40 YR Life Threatening