FDA Adverse Event
Malfunction
Summary report: N
PARADIGM REAL-TIME INSULIN INFUSION PUMP
MDR report key: 3123018
·
Received May 21, 2013
Report
- Report Number
- 3004209178-2013-93814
- Event Type
- Malfunction
- Date Received
- May 21, 2013
- Date of Event
- May 1, 2013
- Report Date
- May 1, 2013
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH
- Reporter Occupation
- PATIENT
Narratives
Additional Manufacturer Narrative · 1
MOTOR ERROR ALARMED DURING REWIND AND ERROR ALARM FOUND IN HISTORY FILE DUE TO CORROSION ON THE MOTOR HOME SWITCH. NO MOISTURE DAMAGE FOUND ON ELECTRONIC ASSEMBLY. ALSO, THE INSULIN PUMP WAS RECEIVED WITH PROTRUDED/LOOSE DRIVE SUPPORT DISK.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE INSULIN PUMP ALARMED MOTOR ERROR. CUSTOMER'S BLOOD GLUCOSE READING IS 46 MG/DL. CUSTOMER TREATED WITH MANUAL INJECTION. THE DRIVE SUPPORT CAP IS PROTRUDED. ADVISED CUSTOMER THE INSULIN PUMP REQUIRES REPLACEMENT. NOTHING FURTHER REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 224578 | PARADIGM REAL-TIME INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-722NAS |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 40 YR | Life Threatening |