FDA Adverse Event Malfunction Summary report: N

HOMECHOICE

MDR report key: 3122988 · Received May 21, 2013

Report

Report Number
1416980-2013-13090
Event Type
Malfunction
Date Received
May 21, 2013
Date of Event
March 29, 2013
Report Date
May 6, 2013
Manufacturer
BAXTER HEALTHCARE - LARGO
Product Code
FKX
PMA / PMN Number
K102936
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS RETURNED TO BAXTER, AND THE EVALUATION IS COMPLETE. (B)(4). THE REPORTED DIFFICULTY OF AN IIPV EVENT WAS CONFIRMED THROUGH AN EVENT HISTORY LOG REVIEW. THE CAUSE WAS UNDETERMINED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A FOLLOW-UP WILL BE SUBMITTED.

Description of Event or Problem · 1

A HIGH DRAIN ERROR 101 (NIGHT DRAIN ONE) ALARM WAS IDENTIFIED IN THE LOG OF A RETURNED HOMECHOICE DEVICE. (A HIGH DRAIN ALARM IS INDICATIVE OF AN IIPV SITUATION. THIS ALARM INDICATES THE PATIENT DRAINED GREATER THAN 200% OF THE MAXIMUM PRESCRIBED CYCLE FILL VOLUME IN STANDARD MODE.) THE ALARM OCCURRED ON (B)(6) 2013 02:47:24. THIS INFORMATION MEETS IIPV CRITERIA. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
225124 HOMECHOICE SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - LARGO

Patients

Seq Age Sex Outcome Treatment
1