FDA Adverse Event Malfunction Summary report: N

ILAB ULTRASOUND IMAGING SYSTEM

MDR report key: 3122932 · Received May 21, 2013

Report

Report Number
2134265-2013-03743
Event Type
Malfunction
Date Received
May 21, 2013
Date of Event
April 23, 2013
Report Date
April 23, 2013
Manufacturer
BOSTON SCIENTIFIC - FREMONT (CE)
Product Code
OBJ
PMA / PMN Number
K051679
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
ES
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATED BY MFR: THE UNIT WAS RECEIVED WITH A DAMAGE LCD DISPLAY COVER. THE UNIT MET SPECIFICATIONS DURING FUNCTIONAL TEST AND SUCCESSFULLY UNDERWENT A CONTINUOUS BURN-IN OVERNIGHT. THE MANUFACTURING BATCH RECORD REVIEW CONFIRMED THAT THE DEVICE MET ALL MATERIAL, ASSEMBLY AND PERFORMANCE SPECIFICATIONS. THE MOST PROBABLE ROOT CAUSE WAS NOT CONFIRMED. (B)(4).

Additional Manufacturer Narrative · 1

AGE AT TIME OF EVENT: 18 YEARS OF AGE OR OVER. COMPLAINANT NAME: HOSPITAL UNIV. (B)(6). (B)(4).

Description of Event or Problem · 1

SAME AS MDR ID # 2134265-2013-03744 AND 2134265-2013-03745. IT WAS REPORTED THAT DURING INTRAVASCULAR ULTRASOUND, AN AUTOMATIC PULLBACK BUTTON DOESN'T WORK. THE PHYSICIAN USED AN ILAB CART SYSTEM 240V MOTOR DRIVE UNIT IN CONJUNCTION WITH A BSC CORONARY IMAGING CATHETER AND ASSY SLED PULLBACK SINGLE PACK MD5 TO IMAGE AN UNSPECIFIED VESSEL. DURING THE PROCEDURE, THE MOTOR DRIVE UNIT'S AUTOMATIC PULLBACK BUTTON WOULD NOT WORK. THEY COMPLETED THE PROCEDURE USING THE TOUCH CONTROL PANEL AUTOMATIC PULLBACK. NO PATIENT COMPLICATIONS WERE REPORTED.

Description of Event or Problem · 1

SAME AS MDR ID # 2134265-2013-03744 AND 2134265-2013-03745. IT WAS REPORTED THAT DURING INTRAVASCULAR ULTRASOUND, AN AUTOMATIC PULLBACK BUTTON DOESN'T WORK. THE PHYSICIAN USED AN ILAB CART SYSTEM 240V MOTOR DRIVE UNIT IN CONJUNCTION WITH A BSC CORONARY IMAGING CATHETER AND ASSY SLED PULLBACK SINGLE PACK MD5 TO IMAGE AN UNSPECIFIED VESSEL. DURING THE PROCEDURE, THE MOTOR DRIVE UNIT'S AUTOMATIC PULLBACK BUTTON WOULD NOT WORK. THEY COMPLETED THE PROCEDURE USING THE TOUCH CONTROL PANEL AUTOMATIC PULLBACK. NO PATIENT COMPLICATIONS WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
224133 ILAB ULTRASOUND IMAGING SYSTEM CATHETER, ULTRASOUND, INTRAVASCULAR OBJ BOSTON SCIENTIFIC - FREMONT (CE) H749ILAB240INS0

Patients

Seq Age Sex Outcome Treatment
1