FDA Adverse Event Malfunction Summary report: N

ACCU-CHEK ® SPIRIT COMBO

MDR report key: 3122767 · Received May 21, 2013

Report

Report Number
2183996-2013-00901
Event Type
Malfunction
Date Received
May 21, 2013
Date of Event
May 14, 2013
Report Date
July 29, 2013
Manufacturer
ROCHE DIABETES CARE AG
Product Code
LZG
PMA / PMN Number
NA
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Additional Manufacturer Narrative · 1

THIS INCIDENT OCCURRED OUTSIDE THE UNITED STATES. INFORMATION CONTAINED WITHIN THIS REPORT IS ALL THAT IS AVAILABLE AT THIS TIME. IF FURTHER INFORMATION IS OBTAINED, IT WILL BE PROVIDED IN A SUPPLEMENTAL REPORT.

Additional Manufacturer Narrative · 1

CONCLUSION: THE COMPLAINT CANNOT BE VERIFIED DUE TO MISHANDLING OF THE PRODUCT. RESULT THE SOFT COMPONENT OF THE UP AND DOWN BUTTON IS DAMAGED DUE TO HANDLING WITH SHARP OBJECTS. THE BUTTONS SHOULDN'T BE ACTUATED BY USING HARD OR SHARP OBJECTS. DUE TO THE LEAKY SOFT COMPONENTS LIQUID ENTERED THE PUMP ELECTRONICS. THE RESULTING SHORT CIRCUIT DAMAGED THE PUMP ELECTRONICS AND LED TO THE FOLLOWING E8 ERROR MESSAGE(S): E8 WERE FOUND IN THE HISTORY. THE UP AND DOWN BUTTON IS ALWAYS ACTIVE. DUE TO AN EXTERNAL MECHANICAL INFLUENCE THE SNAP DOME OF THE BUTTONS IS PRESSED THROUGH. THIS SOURCE LED TO AN ALWAYS ACTIVATED UP AND DOWN BUTTON AND UNCLEAR ABLE E8 ERROR MESSAGES. THE PROBLEM MENTIONED BY THE CUSTOMER IS RELATED TO MISHANDLING OF THE PRODUCT BY THE CUSTOMER.

Description of Event or Problem · 1

PATIENT REPORTED THAT ON (B)(6) 2013, THE INFUSION DEVICE WAS OFF AND NOT RESPONDING. PATIENT STATED THE INFUSION DEVICE DID NOT DISPLAY AN ALARM OR ERROR MESSAGE. PATIENT REPORTED THE BLOOD GLUCOSE LEVEL IS OKAY; NO EXACT READING WAS PROVIDED. PATIENT'S NORMAL BLOOD GLUCOSE RANGE WAS NOT PROVIDED. NO FURTHER INFORMATION IS AVAILABLE. NO PHYSIOLOGICAL EFFECTS WERE REPORTED. THE PATIENT DID NOT REQUIRE ASSISTANCE FROM A HEALTHCARE PROFESSIONAL OR SECOND PARTY TO ADDRESS THE ISSUE. REQUESTED RETURN OF THE ALLEGED INFUSION DEVICE FOR EVALUATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
224885 ACCU-CHEK ® SPIRIT COMBO INSULIN INFUSION PUMP LZG ROCHE DIABETES CARE AG 00700006863 NA

Patients

Seq Age Sex Outcome Treatment
1