ACCU-CHEK ® SPIRIT COMBO
Report
- Report Number
- 2183996-2013-00901
- Event Type
- Malfunction
- Date Received
- May 21, 2013
- Date of Event
- May 14, 2013
- Report Date
- July 29, 2013
- Manufacturer
- ROCHE DIABETES CARE AG
- Product Code
- LZG
- PMA / PMN Number
- NA
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE
Narratives
THIS INCIDENT OCCURRED OUTSIDE THE UNITED STATES. INFORMATION CONTAINED WITHIN THIS REPORT IS ALL THAT IS AVAILABLE AT THIS TIME. IF FURTHER INFORMATION IS OBTAINED, IT WILL BE PROVIDED IN A SUPPLEMENTAL REPORT.
CONCLUSION: THE COMPLAINT CANNOT BE VERIFIED DUE TO MISHANDLING OF THE PRODUCT. RESULT THE SOFT COMPONENT OF THE UP AND DOWN BUTTON IS DAMAGED DUE TO HANDLING WITH SHARP OBJECTS. THE BUTTONS SHOULDN'T BE ACTUATED BY USING HARD OR SHARP OBJECTS. DUE TO THE LEAKY SOFT COMPONENTS LIQUID ENTERED THE PUMP ELECTRONICS. THE RESULTING SHORT CIRCUIT DAMAGED THE PUMP ELECTRONICS AND LED TO THE FOLLOWING E8 ERROR MESSAGE(S): E8 WERE FOUND IN THE HISTORY. THE UP AND DOWN BUTTON IS ALWAYS ACTIVE. DUE TO AN EXTERNAL MECHANICAL INFLUENCE THE SNAP DOME OF THE BUTTONS IS PRESSED THROUGH. THIS SOURCE LED TO AN ALWAYS ACTIVATED UP AND DOWN BUTTON AND UNCLEAR ABLE E8 ERROR MESSAGES. THE PROBLEM MENTIONED BY THE CUSTOMER IS RELATED TO MISHANDLING OF THE PRODUCT BY THE CUSTOMER.
PATIENT REPORTED THAT ON (B)(6) 2013, THE INFUSION DEVICE WAS OFF AND NOT RESPONDING. PATIENT STATED THE INFUSION DEVICE DID NOT DISPLAY AN ALARM OR ERROR MESSAGE. PATIENT REPORTED THE BLOOD GLUCOSE LEVEL IS OKAY; NO EXACT READING WAS PROVIDED. PATIENT'S NORMAL BLOOD GLUCOSE RANGE WAS NOT PROVIDED. NO FURTHER INFORMATION IS AVAILABLE. NO PHYSIOLOGICAL EFFECTS WERE REPORTED. THE PATIENT DID NOT REQUIRE ASSISTANCE FROM A HEALTHCARE PROFESSIONAL OR SECOND PARTY TO ADDRESS THE ISSUE. REQUESTED RETURN OF THE ALLEGED INFUSION DEVICE FOR EVALUATION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 224885 | ACCU-CHEK ® SPIRIT COMBO | INSULIN INFUSION PUMP | LZG | ROCHE DIABETES CARE AG | 00700006863 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |