FDA Adverse Event Malfunction Summary report: N

PROXIMATE ILS CURVED CIR STAPL

MDR report key: 3122742 · Received May 21, 2013

Report

Report Number
3005075853-2013-02472
Event Type
Malfunction
Date Received
May 21, 2013
Date of Event
April 3, 2013
Report Date
May 1, 2013
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
GDW
PMA / PMN Number
K983536
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). INFORMATION WAS NOT PROVIDED BY CONTACT. INFORMATION ANTICIPATED, BUT UNAVAILABLE AT THIS TIME.

Additional Manufacturer Narrative · 1

(B)(4). THE ANALYSIS RESULTS FOUND THAT THE DEVICE ARRIVED IN GOOD VISUAL CONDITION. THE BREAKAWAY WASHER WAS PRESENT AND CUT AND THERE WERE NO STAPLES PRESENT, INDICATING THAT THE DEVICE ACHIEVED A FULL FIRING STROKE. THE DEVICE WAS RELOADED WITH STAPLES, A NEW WASHER WAS PLACED ON THE DEVICE AND IT WAS TESTED FOR FUNCTIONALITY. IT FIRED AND FORMED ALL THE STAPLES AS WELL AS COMPLETELY CUT THE TEST MEDIA AND THE BREAKAWAY WASHER WITHOUT INCIDENT. THE STAPLE LINE WAS COMPLETE AND THE STAPLES WERE NOTED TO HAVE THE PROPER B-FORMED SHAPE. THE MANUFACTURING RECORDS WERE REVIEWED AND NO DISCREPANCIES WERE FOUND DURING THE MANUFACTURING PROCESS.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING AN ANTERIOR RESECTION PROCEDURE THE CONSULTANT HAD TROUBLE REMOVING THE SPIKE FROM THE HEAD OF THE DEVICE. IT ADDED 10-15 MINUTES ONTO THE PROCEDURE TIME. HOWEVER, THEY WERE ABLE TO REMOVE AND FIRE THE DEVICE FULLY WITH NO OTHER PROBLEMS. NO PATIENT CONSEQUENCES REPORTED.