FDA Adverse Event Malfunction Summary report: N

COATED VICRYL PLUS ANTIBACTERIAL (POLYGLACTIN 910)

MDR report key: 3122735 · Received May 21, 2013

Report

Report Number
2210968-2013-05751
Event Type
Malfunction
Date Received
May 21, 2013
Date of Event
April 29, 2013
Report Date
April 29, 2013
Manufacturer
ETHICON, INC.
Product Code
GAM
PMA / PMN Number
K032420
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KS, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4). REPRESENTATIVE SAMPLES WERE RETURNED FOR EVALUATION. THEY WERE VISUALLY AND FUNCTIONALLY EXAMINED FOR TENSILE STRENGTH AND THEY MET THE REQUIREMENTS. IN ADDITION, A REVIEW OF THE BATCH MANUFACTURING RECORDS WAS CONDUCTED AND THE BATCH MET ALL FINISHED GOODS RELEASE CRITERIA. THIS IS ONE OF FIVE MEDWATCHES BEING SUBMITTED AS FIVE DEVICES WERE INVOLVED IN THIS EVENT. SEE ALSO MEDWATCHES 2210968-2013-05750, 2210968-2013-05752, 2210968-2013-05753, AND 2210968-2013-05754. THE SAME PATIENT IS REPRESENTED IN EACH MEDWATCH.

Description of Event or Problem · 1

IT WAS REPORTED THAT A PATIENT UNDERWENT A SPINE PROCEDURE ON (B)(6) 2013 AND SUTURE WAS USED. DURING THE PROCEDURE, THE NEEDLE PULLED OFF THE SUTURE. THE PROCEDURE WAS COMPLETED WITH A LIKE DEVICE. THERE WERE NO ADVERSE PATIENT CONSEQUENCES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
224877 COATED VICRYL PLUS ANTIBACTERIAL (POLYGLACTIN 910) SUTURE, ABSORBABLE GAM ETHICON, INC. NA EM6178

Patients

Seq Age Sex Outcome Treatment
1