FDA Adverse Event Malfunction Summary report: N

ENSEAL G2 CURVED JAW

MDR report key: 3122688 · Received May 21, 2013

Report

Report Number
3005075853-2013-02467
Event Type
Malfunction
Date Received
May 21, 2013
Report Date
April 30, 2013
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
GEI
PMA / PMN Number
K112033
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS RECEIVED WITH THE I-BLADE OUT OF THE BUTTON JAW AND THE ELECTRODE SEPARATED. THE DEVICE WAS PARTIALLY TESTED WITH A GENERATOR AND THE DEVICE PERFORMED AS REQUIRED DURING TESTING; THOUGH ALL TESTING COULD NOT BE COMPLETED DUE TO THE DAMAGED I BLADE. THE CONDITION OF THE BLADE PREVENTED THE FUNCTIONALITY OF THE JAW. THE JAW WAS UNABLE TO OPEN AND CLOSE. IT IS POSSIBLE THAT A REPLACE INSTRUMENT WAS NOTED DURING USAGE WITH THE ELECTRODE SEPARATED IN THIS MANNER. THIS WAS NOT CONFIRMED DURING COMPLAINT ANALYSIS. THERE IS INSUFFICIENT EVIDENCE RELATED TO WHAT CAUSED THE DAMAGE; A PROBABLE CAUSE OF THE DAMAGE TO THE I-BLADE COULD BE NOT ALLOWING THICK AND FIBROUS TISSUES TO DENATURE PRIOR TO I-BLADE ADVANCEMENT.

Additional Manufacturer Narrative · 1

(B)(4). ADDITIONAL INFORMATION: AFTER SEVERAL REQUESTS, THE DEVICE WAS NOT RECEIVED FOR ANALYSIS. SHOULD THE DEVICE BE RETURNED FOR ANALYSIS, A SUPPLEMENTAL MEDWATCH WILL BE SENT

Additional Manufacturer Narrative · 1

(B)(4). WHEN ADDITIONAL INFORMATION IS RECEIVED AND/OR THE DEVICE ANALYSIS HAS BEEN COMPLETED, A SUPPLEMENTAL MEDWATCH WILL BE SENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A LAPAROSCOPIC HYSTERECTOMY PROCEDURE THERE WAS AN ERROR CODE: ¿REPLACE INSTRUMENT¿ AND BLADE WAS DEFORMED; WITH GEN11. NO FURTHER INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
223943 ENSEAL G2 CURVED JAW ELECTROSURGICAL CUTTING & COAGULATION & ACCESSORIES GEI ETHICON ENDO-SURGERY, LLC. UNK K4CD6E

Patients

Seq Age Sex Outcome Treatment
1 GENERATOR