ENSEAL G2 CURVED JAW
Report
- Report Number
- 3005075853-2013-02467
- Event Type
- Malfunction
- Date Received
- May 21, 2013
- Report Date
- April 30, 2013
- Manufacturer
- ETHICON ENDO-SURGERY, LLC.
- Product Code
- GEI
- PMA / PMN Number
- K112033
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- OTHER
Narratives
(B)(4). THE DEVICE WAS RECEIVED WITH THE I-BLADE OUT OF THE BUTTON JAW AND THE ELECTRODE SEPARATED. THE DEVICE WAS PARTIALLY TESTED WITH A GENERATOR AND THE DEVICE PERFORMED AS REQUIRED DURING TESTING; THOUGH ALL TESTING COULD NOT BE COMPLETED DUE TO THE DAMAGED I BLADE. THE CONDITION OF THE BLADE PREVENTED THE FUNCTIONALITY OF THE JAW. THE JAW WAS UNABLE TO OPEN AND CLOSE. IT IS POSSIBLE THAT A REPLACE INSTRUMENT WAS NOTED DURING USAGE WITH THE ELECTRODE SEPARATED IN THIS MANNER. THIS WAS NOT CONFIRMED DURING COMPLAINT ANALYSIS. THERE IS INSUFFICIENT EVIDENCE RELATED TO WHAT CAUSED THE DAMAGE; A PROBABLE CAUSE OF THE DAMAGE TO THE I-BLADE COULD BE NOT ALLOWING THICK AND FIBROUS TISSUES TO DENATURE PRIOR TO I-BLADE ADVANCEMENT.
(B)(4). ADDITIONAL INFORMATION: AFTER SEVERAL REQUESTS, THE DEVICE WAS NOT RECEIVED FOR ANALYSIS. SHOULD THE DEVICE BE RETURNED FOR ANALYSIS, A SUPPLEMENTAL MEDWATCH WILL BE SENT
(B)(4). WHEN ADDITIONAL INFORMATION IS RECEIVED AND/OR THE DEVICE ANALYSIS HAS BEEN COMPLETED, A SUPPLEMENTAL MEDWATCH WILL BE SENT.
IT WAS REPORTED THAT DURING A LAPAROSCOPIC HYSTERECTOMY PROCEDURE THERE WAS AN ERROR CODE: ¿REPLACE INSTRUMENT¿ AND BLADE WAS DEFORMED; WITH GEN11. NO FURTHER INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 223943 | ENSEAL G2 CURVED JAW | ELECTROSURGICAL CUTTING & COAGULATION & ACCESSORIES | GEI | ETHICON ENDO-SURGERY, LLC. | UNK | K4CD6E |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | GENERATOR |