DUAL TRIGGER ROTARY
Report
- Report Number
- 0001811755-2013-01160
- Event Type
- Malfunction
- Date Received
- May 21, 2013
- Date of Event
- May 13, 2013
- Report Date
- May 13, 2013
- Manufacturer
- STRYKER INSTRUMENTS-KALAMAZOO
- Product Code
- KIJ
- PMA / PMN Number
- K972367
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WA, US
- Reporter Occupation
- OTHER
Narratives
THE DEVICE HAS NOT BEEN RETURNED FOR ANALYSIS AT THIS TIME.
THE REPORTED RUN-ON WAS CONFIRMED BY A MANUFACTURER SERVICE TECHNICIAN THROUGH FUNCTIONAL EVALUATION. A DAMAGED TRIGGER WAS IDENTIFIED, WHICH IS BELIEVED TO BE THE PROBABLE CAUSE OF THE REPORTED EVENT.
IT WAS REPORTED THAT DURING EQUIPMENT TESTING CONDUCTED AT THE USER FACILITY THE DEVICE WAS RUNNING WITHOUT USER ACTIVATION. NO MEDICAL INTERVENTION AND NO ADVERSE CONSEQUENCES WERE ALLEGED WITH THIS EVENT. AS THIS EVENT OCCURRED DURING TESTING, THERE WAS NO PATIENT INVOLVEMENT AND NO DELAY TO A SURGICAL PROCEDURE.
IT WAS REPORTED THAT DURING EQUIPMENT TESTING CONDUCTED AT THE USER FACILITY THE DEVICE WAS RUNNING WITHOUT USER ACTIVATION. NO MEDICAL INTERVENTION AND NO ADVERSE CONSEQUENCES WERE ALLEGED WITH THIS EVENT. AS THIS EVENT OCCURRED DURING TESTING, THERE WAS NO PATIENT INVOLVEMENT AND NO DELAY TO A SURGICAL PROCEDURE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 224497 | DUAL TRIGGER ROTARY | INSTRUMENT, SURGICAL, ORTHOPEDIC, DC-POWERED MOTOR AND ACCESSORY/ATTACHMENT | KIJ | STRYKER INSTRUMENTS-KALAMAZOO |