FDA Adverse Event Malfunction Summary report: N

DUAL TRIGGER ROTARY

MDR report key: 3122686 · Received May 21, 2013

Report

Report Number
0001811755-2013-01160
Event Type
Malfunction
Date Received
May 21, 2013
Date of Event
May 13, 2013
Report Date
May 13, 2013
Manufacturer
STRYKER INSTRUMENTS-KALAMAZOO
Product Code
KIJ
PMA / PMN Number
K972367
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE HAS NOT BEEN RETURNED FOR ANALYSIS AT THIS TIME.

Additional Manufacturer Narrative · 1

THE REPORTED RUN-ON WAS CONFIRMED BY A MANUFACTURER SERVICE TECHNICIAN THROUGH FUNCTIONAL EVALUATION. A DAMAGED TRIGGER WAS IDENTIFIED, WHICH IS BELIEVED TO BE THE PROBABLE CAUSE OF THE REPORTED EVENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING EQUIPMENT TESTING CONDUCTED AT THE USER FACILITY THE DEVICE WAS RUNNING WITHOUT USER ACTIVATION. NO MEDICAL INTERVENTION AND NO ADVERSE CONSEQUENCES WERE ALLEGED WITH THIS EVENT. AS THIS EVENT OCCURRED DURING TESTING, THERE WAS NO PATIENT INVOLVEMENT AND NO DELAY TO A SURGICAL PROCEDURE.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING EQUIPMENT TESTING CONDUCTED AT THE USER FACILITY THE DEVICE WAS RUNNING WITHOUT USER ACTIVATION. NO MEDICAL INTERVENTION AND NO ADVERSE CONSEQUENCES WERE ALLEGED WITH THIS EVENT. AS THIS EVENT OCCURRED DURING TESTING, THERE WAS NO PATIENT INVOLVEMENT AND NO DELAY TO A SURGICAL PROCEDURE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
224497 DUAL TRIGGER ROTARY INSTRUMENT, SURGICAL, ORTHOPEDIC, DC-POWERED MOTOR AND ACCESSORY/ATTACHMENT KIJ STRYKER INSTRUMENTS-KALAMAZOO

Patients

Seq Age Sex Outcome Treatment
1