Narratives
PRODUCT EVENT SUMMARY: (B)(4). THE PROXIMAL SEGMENT OF THE LEAD WAS RETURNED, ANALYZED AND NO ANOMALIES WERE FOUND, THERE WAS BLOOD ON THE PROXIMAL CONDUCTOR OF THE LEAD AND IT WAS NOT OBSTRUCTED, THERE WAS BLOOD ON THE DISTAL CONDUCTOR OF THE LEAD AND IT WAS NOT OBSTRUCTED, THE OUTER INSULATION OF THE LEAD DEVELOPED A COSMETIC DEPRESSION WHILE IN VIVO, VISUAL SUMMARY ANALYSIS OF THE LEAD WAS PERFORMED ONLY.
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. THE DEVICE(S) ASSOCIATED WITH THIS ADVERSE OUTCOME WAS/WERE RETURNED FROM AN UNKNOWN SOURCE GREATER THAN TWO YEARS FROM TODAY. NO CONTACTS/EVENTS REGARDING THE SYSTEM HAVE EVER BEEN RECEIVED/REPORTED. CONSEQUENTLY, CONTACT INFORMATION TO COMPLETE FOLLOW-UP IS NOT REASONABLY KNOWN. THEREFORE, ATTEMPTS FOR ADDITIONAL INFORMATION CANNOT BE MADE. (B)(4) IMPLANTED: (B)(6) 2010; PRODUCT ID 5076-52 IMPLANTED: (B)(6) 2010. (B)(4).
IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
AN IMPLANTABLE PULSE GENERATOR (IPG) SYSTEM WAS RETURNED FROM AN UNKNOWN SOURCE WITH NO INFORMATION. INFORMATION IDENTIFIED IN THE MANUFACTURER¿S DATABASE INDICATED THE PATIENT DIED APPROXIMATELY TEN MONTHS POST IMPLANT OF THE IPG SYSTEM APPROXIMATELY TWO YEARS AGO.