FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 3122662 · Received May 21, 2013

Report

Report Number
2531779-2013-06825
Event Type
Malfunction
Date Received
May 21, 2013
Report Date
April 27, 2013
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.

Additional Manufacturer Narrative · 1

FOLLOW-UP #1 08/08/2013 ¿ DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2013 WITH THE FOLLOWING FINDINGS: THE PUMP BLACK BOX SHOWED A ¿PARTIAL DELIVERY, USER ABORTED (PUMP)¿ WARNING OCCURRED ON 04/26/2013. DURING TESTING A NORMAL BOLUS AND AN AUDIO BOLUS WERE PROGRAMMED AND DELIVERED SUCCESSFULLY AND WERE RECORDED ACCURATELY IN THE PUMP HISTORY. THE KEYPAD WAS TESTED AND CONFIRMED THAT THERE WERE NO HYPERSENSITIVE KEYS. THE PUMP WAS EXERCISED FOR 24 HOURS WITHOUT ANY ERRORS OR WARNINGS OCCURRING. THE COMPLAINT WAS UNABLE TO BE DUPLICATED DURING TESTING.

Description of Event or Problem · 1

ON (B)(6) 2013, THE REPORTER CONTACTED ANIMAS, ALLEGING A HISTORY/SETTINGS (HISTORY/SETTINGS ISSUE) ISSUE. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.