FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 3122644 · Received May 21, 2013

Report

Report Number
2531779-2013-06823
Event Type
Malfunction
Date Received
May 21, 2013
Report Date
April 25, 2013
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

FOLLOW-UP # 1 DATE OF SUBMISSION 07/17/2013 - DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 06/18/2013 WITH THE FOLLOWING FINDINGS: DURING TESTING, THE COMPLAINT OF A BLANK DISPLAY SCREEN WAS NOT DUPLICATED. THE DISPLAY FUNCTIONED PROPERLY AND WAS FULLY ILLUMINATED WITH NO VISIBLE SIGNS OF FADING OR DISCOLORATION. UNRELATED TO THE COMPLAINT, EVALUATION REVEALED THAT THE BATTERY COMPARTMENT THREADS WERE DAMAGED. THE BATTERY CAP WAS NOT ABLE TO FULLY TIGHTEN DUE TO DAMAGED CAP THREADS AND DAMAGED BATTERY CAP CONTACTS.

Additional Manufacturer Narrative · 1

THE PUMP HAS BEEN RETURNED TO ANIMAS BUT EVALUATION HAS NOT YET BEEN COMPLETED. WHEN EVALUATION IS COMPLETE A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSION CAN BE MADE AT THIS TIME.

Description of Event or Problem · 1

ON (B)(6) 2013, THE REPORTER CONTACTED ANIMAS, ALLEGING A DISPLAY (BLANK SCREEN) ISSUE. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
224427 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 40 YR