FDA Adverse Event
Malfunction
Summary report: N
LIGACLIP** ENDOSCOPIC CLIP APPLIER
MDR report key: 3122571
·
Received May 21, 2013
Report
- Report Number
- 3005075853-2013-02464
- Event Type
- Malfunction
- Date Received
- May 21, 2013
- Date of Event
- March 5, 2013
- Report Date
- April 26, 2013
- Manufacturer
- ETHICON ENDO-SURGERY, LLC.
- Product Code
- FZP
- PMA / PMN Number
- K864102
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TU
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
(B)(4). INFORMATION WAS NOT PROVIDED BY THE INITIAL CONTACT. THE ANALYSIS RESULTS FOUND THAT THE DEVICE WAS RETURNED IN GOOD VISUAL CONDITION. IN AN ATTEMPT TO REPLICATE THE REPORTED INCIDENT, THE DEVICE WAS TESTED FOR FUNCTIONALITY. UPON TESTING, THE DEVICE WAS CYCLED, FED, RETAINED AND FORMED THE REMAINING CLIPS AS INTENDED. THE DEVICE WAS FOUND TO BE FULLY FUNCTIONAL AND CONFORMING TO OUR MANUFACTURING SPECIFICATIONS. NO CONCLUSION COULD BE REACHED AS TO WHAT MAY HAVE CAUSED THE REPORTED INCIDENT. THE BATCH RECORD WAS REVIEWED AND NO ANOMALIES WERE NOTED DURING THE MANUFACTURING PROCESS.
Description of Event or Problem · 1
IT WAS REPORTED THAT DURING A LAPAROSCOPIC SURGERY, THE CLIPS DID NOT WORK PROPERLY AND THE TISSUE WAS DAMAGED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 225045 | LIGACLIP** ENDOSCOPIC CLIP APPLIER | CLIP, IMPLANTABLE | FZP | ETHICON ENDO-SURGERY, LLC. | NA | J4CV4U |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |