FDA Adverse Event Injury Summary report: N

OLYMPUS ELECTROSURGICAL UNIT

MDR report key: 3122563 · Received May 17, 2013

Report

Report Number
9610773-2013-00035
Event Type
Injury
Date Received
May 17, 2013
Date of Event
April 18, 2013
Report Date
April 23, 2013
Manufacturer
OLYMPUS WINTER & IBE GMBH
Product Code
KNS
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
ND, US
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

OLYMPUS FOLLOWED UP WITH THE USER FACILITY TO OBTAIN ADD'L INFO REGARDING THE REPORT AND WAS INFORMED THAT THE PROCEDURE WAS A DIAGNOSTIC COLONOSCOPY. THE PT WAS SEDATED WITH PROPOFOL. DURING THE EXAM, A 10 MM POLYP WAS NOTED IN THE CECUM. THE USER UTILIZED THE SUBJECT DEVICE VIA THE YELLOW FOOT PEDAL WITH A COOK SOFT ACUSNARE TO CAUTERIZE THE POLYP. IT WAS REPORTED THAT THERE WERE SOME REMNANTS OF THE POLYP LEFT. HENCE THE USER ACTIVATED THE YELLOW PEDAL USED WITH THE SUBJECT DEVICE AGAIN TO CAUTERIZE THE EDGE OF THE POLYP SITE AND THE PT REPORTEDLY JOLTED. THERE WAS NO REPORT OF PT INJURY. THE SETTINGS ON THE SUBJECT DEVICE WERE MONOPOLAR PULSECUT SLOW 120. THE PT WAS SENT HOME THE SAME DAY. THE FOLLOWING DAY THE PT REPORTEDLY COMPLAINED OF ABDOMINAL PAIN AND FEVER, AND WAS ADMITTED TO THE HOSPITAL. FURTHER EXAMINATION OF THE PT REVEALED THAT THERE WAS NO EVIDENCE OF BOWEL PERFORATION OR OTHER INJURIES. THE PT WAS DISCHARGED 24 HOURS LATER AND IS REPORTEDLY DORING FINE. THE SUBJECT DEVICE WAS TESTED BY THE USER FACILITY'S BIOMED DEPARTMENT AND IT WAS REPORTED TO BE WORKING PROPERLY. IN ADDITION, THE SUBJECT DEVICE WAS RETURNED TO OLYMPUS FOR EVAL. THE DEVICE PASSED ALL THE ELECTRICAL SAFETY TESTS. IN ADDITION, THE USER FACILITY'S VALLEYLAB SPLIT PT PAD WAS TESTED WITH THE SUBJECT DEVICE AND IT WAS WORKING PROPERLY. THE USER'S EXPERIENCE CANNOT BE CONCLUSIVELY DETERMINED AT THIS TIME.

Description of Event or Problem · 1

OLYMPUS WAS INFORMED THAT DURING AN UNSPECIFIED PROCEDURE "THE PT JOLTED AS IF BEING SHOCKED." THERE WAS REPORT OF PT INJURY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
218821 OLYMPUS ELECTROSURGICAL UNIT ELECTROSURGICAL UNIT KNS OLYMPUS WINTER & IBE GMBH ESG-100 (WB991046) NA

Patients

Seq Age Sex Outcome Treatment
1 56 YR Other COOK SOFT ACUSNARE: MODEL G22632, LOT # UNK| OLYMPUS CF-HQ190L: LOT # UNK| VALLEYLAB SPLIT PAD: LOT # 251069X