ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2013-06817
- Event Type
- Injury
- Date Received
- May 21, 2013
- Report Date
- April 25, 2013
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER
Narratives
THE PUMP HAS BEEN RETURNED TO ANIMAS. EVALUATION HAS NOT YET BEEN COMPLETED. WHEN EVALUATION IS COMPLETE A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSION CAN BE MADE AT THIS TIME. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.
FOLLOW-UP # 1 (B)(4) - DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2013 WITH THE FOLLOWING FINDINGS: DURING INVESTIGATION, THE PUMP¿S BLACK BOX HISTORY WAS REVIEWED AND SHOWED THE LAST BASAL DELIVERY AND THE LAST BOLUS WERE RECORDED ON (B)(6) 2013. THERE WERE NO ALARMS RELATED TO THE COMPLAINT IN THE BLACK BOX OR ALARM HISTORY; ONLY TYPICAL USAGE ALARMS AND WARNINGS WERE OBSERVED. THE TOTAL DAILY DOSAGE¿S ADDED UP CORRECTLY TO REFLECT THE USER¿S PROGRAMMED BASAL RATES. THE PUMP SUCCESSFULLY COMPLETED THE REQUIRED (B)(4) FLOW ACCURACY TEST AND WAS FOUND TO BE DELIVERING WITHIN THE REQUIRED SPECIFICATIONS. THE COMPLAINT COULD NOT BE DUPLICATED DURING THE INVESTIGATION AND THERE WAS NO DEFECT FOUND.
ON (B)(6) 2013, THE REPORTER CONTACTED ANIMAS AND ALLEGED THAT WHILE ON AN ALTERNATE DELIVERY METHOD, WAITING FOR A REPLACEMENT PUMP TO ARRIVE, SHE FAINTED AT WORK , WAS TAKEN TO URGENT CARE AND TRANSPORTED TO THE HOSPITAL. SHE HAD HYPOGLYCEMIC EXCURSIONS, THE LOWEST BG WAS 29 MG/DL. THIS COMPLAINT IS BEING REPORTED BECAUSE THE PATIENT EXPERIENCED HYPOGLYCEMIA AFTER BEING TOLD BY CUSTOMER SUPPORT (CS) TO DISCONTINUE USING THEIR PUMP.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 223759 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 41 YR | Hospitalization |