FDA Adverse Event Malfunction Summary report: N

NC QUANTUM APEX?

MDR report key: 3122464 · Received May 21, 2013

Report

Report Number
2134265-2013-03312
Event Type
Malfunction
Date Received
May 21, 2013
Date of Event
April 8, 2013
Report Date
April 23, 2013
Manufacturer
BOSTON SCIENTIFIC - MAPLE GROVE
Product Code
LOX
PMA / PMN Number
P860019
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4).

Additional Manufacturer Narrative · 1

DEVICE EVALUATED BY MFR: THE NC QUANTUM APEX CATHETER WAS RECEIVED INSIDE A NC QUANTUM APEX SHELF BOX THAT MATCHED THE INFORMATION ON THE DEVICE. THERE WAS BLOOD IN THE BALLOON AND INFLATION LUMEN. MAGNIFIED INSPECTION REVEALED A LONGITUDINAL TEAR IN THE BALLOON WALL FROM THE MID-SECTION TO DISTAL END OF THE BALLOON. THE BALLOON PRESENTED NO IRREGULARITIES IN THE BALLOON MATERIAL OR THE RO MARKER THAT COULD HAVE CONTRIBUTED TO THE DAMAGE. THERE WAS NO EVIDENCE OF ANY PRODUCT QUALITY DEFICIENCIES CONTRIBUTING TO THE DAMAGE. THE MANUFACTURING BATCH RECORD REVIEW CONFIRMED THAT THE DEVICE MET ALL MATERIAL, ASSEMBLY AND PERFORMANCE SPECIFICATIONS. THE MOST PROBABLE ROOT CAUSE IS OPERATIONAL CONTEXT AS DEVICE PERFORMANCE WAS LIMITED DUE TO ANATOMICAL PROCEDURAL FACTORS. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING PERCUTANEOUS CORONARY INTERVENTION, A BALLOON RUPTURE OCCURRED. THE 12MM X 2.50MM NC QUANTUM APEX BALLOON CATHETER RUPTURED AT APPROXIMATELY RATED BURST PRESSURE. THE BALLOON WAS REMOVED INTACT. THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED AND THE PATIENT STATUS IS GOOD.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING PERCUTANEOUS CORONARY INTERVENTION, A BALLOON RUPTURE OCCURRED. THE 12MM X 2.50MM NC QUANTUM APEX BALLOON CATHETER RUPTURED AT APPROXIMATELY RATED BURST PRESSURE. THE BALLOON WAS REMOVED INTACT. THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED AND THE PATIENT STATUS IS GOOD.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
223625 NC QUANTUM APEX? CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS LOX BOSTON SCIENTIFIC - MAPLE GROVE H7493912412250 14593462

Patients

Seq Age Sex Outcome Treatment
1