COULTER® LH 750 HEMATOLOGY ANALYZER
Report
- Report Number
- 1061932-2013-00909
- Event Type
- Malfunction
- Date Received
- May 21, 2013
- Date of Event
- April 23, 2013
- Report Date
- April 23, 2013
- Manufacturer
- BECKMAN COULTER
- Product Code
- GKZ
- PMA / PMN Number
- K061574
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NM, US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
A BEC FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED ON (B)(4) 2013 FOR THIS EVENT. THE FSE INDICATED THAT THE INSTRUMENT WAS DRIPPING A FEW DROPS OF BACKWASH DILUENT FROM THE SECONDARY ASPIRATION PROBE WHEN THE INSTRUMENTS COMPRESSOR SHUTS DOWN DUE TO A PROGRAMMED SHUTDOWN OR WHEN THE IDLE TIMER TIMES OUT AND SHUTS DOWN THE COMPRESSOR. THE FSE FOUND THE NEEDLE BACKWASH LINE WAS CAUSING THE PROBLEM. THE FSE REPLACED THE NEEDLE BACKWASH DILUENT PUMP AND THE ASSOCIATED TUBING FROM THE BACKWASH MANIFOLD 4 TO PINCH VALVE (PV) 59. THE FSE ALSO FOUND THAT THE LIGHT SCATTER OFFSET VOLTAGE WAS TOO HIGH ON STARTUP. THE FSE THEN CLEANED THE FLOWCELL AND BROUGHT THE BACKGROUND OFFSET VOLTAGE BACK INTO SPECIFICATION. FAILURE MODE: NEEDLE BACKWASH DILUENT PUMP AND ASSOCIATED TUBING WHICH NEEDED TO BE REPLACED. PER LABELING, BECKMAN COULTER URGES ITS CUSTOMERS TO COMPLY WITH ALL NATIONAL HEALTH AND SAFETY STANDARDS SUCH AS THE USE OF BARRIER PROTECTION. THIS MAY INCLUDE, BUT IS NOT LIMITED TO PROTECTIVE EYEWEAR, GLOVES AND SUITABLE LABORATORY ATTIRE WHEN OPERATING OR MAINTAINING THIS OR ANY OTHER AUTOMATED LABORATORY ANALYZER. (B)(4).
A CUSTOMER CONTACTED BECKMAN COULTER (BEC) REPORTING THAT THE PROBE IN THE COULTER LH 750 HEMATOLOGY ANALYZER STATION HAS BEEN LEAKING CLEAR FLUID SINCE THE PREVENTIVE MAINTENANCE (PM) PROCEDURE WAS PERFORMED LAST WEEK. THE VOLUME OF THE LEAK WAS LESS THAN 1 ML (ABOUT THREE DROPS) AND WAS NOT CONTAINED WITHIN THE INSTRUMENT. THE OPERATOR WAS WEARING PERSONAL PROTECTIVE EQUIPMENT (PPE) CONSISTING OF GLOVES AND A LABORATORY COAT AT THE TIME OF THE INCIDENT. THE MATERIAL SAFETY DATA SHEET (MSDS) WAS NOT REVIEWED. THERE IS AN EXPOSURE CONTROL / RISK MANAGEMENT PLAN IN PLACE AT THE FACILITY. THERE WAS NO BIOHAZARD EXPOSURE TO MUCOUS MEMBRANES OR OPEN WOUNDS. MEDICAL ATTENTION WAS NOT SOUGHT. NO ERRONEOUS RESULTS WERE GENERATED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 224644 | COULTER® LH 750 HEMATOLOGY ANALYZER | COUNTER, DIFFERENTIAL CELL | GKZ | BECKMAN COULTER | LH 750 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |