FDA Adverse Event Malfunction Summary report: N

PROXIMATE PPH PROCEDURE SET

MDR report key: 3122400 · Received May 21, 2013

Report

Report Number
3005075853-2013-02460
Event Type
Malfunction
Date Received
May 21, 2013
Date of Event
January 11, 2013
Report Date
April 29, 2013
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
GDW
PMA / PMN Number
K051301
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). BREAKAWAY WASHER CUT OFF CENTER. THE ANALYSIS RESULTS FOUND THAT THE DEVICE ARRIVED WITH THE KNIFE DAMAGED; THE BREAKAWAY WASHER WAS PRESENT AND WITH AN OFF-CENTER CUT. IT APPEARS POSSIBLE THAT THE ANVIL WAS PUSHED FAR ENOUGH OFF CENTER TO RESULT IN AN OFF CENTER CUT OF THE BREAKAWAY WASHER AND DAMAGE THE KNIFE BY PRESSING IT HARD ENOUGH AGAINST THE ANVIL. THIS SITUATION NORMALLY OCCURS WHEN THE TISSUE IS NOT EVENLY DISTRIBUTED IN THE DEVICE. THE DEVICE WAS RELOADED WITH STAPLES, A NEW WASHER WAS PLACED AND THE DEVICE WAS TESTED FOR FUNCTIONALITY, IT FIRED AND FORMED ALL THE STAPLES AS WELL AS COMPLETELY CUT THE TEST MEDIA AND THE BREAKAWAY WASHER WITHOUT INCIDENT, HOWEVER THE WASHER CUT WAS NOT A PERFECT CIRCLE DUE TO THE DAMAGED KNIFE. THE STAPLE LINE WAS COMPLETE AND THE STAPLES WERE NOTED TO HAVE THE PROPER B-FORMED SHAPE. THE MANUFACTURING RECORDS WERE REVIEWED AND NO DISCREPANCIES WERE FOUND DURING THE MANUFACTURING PROCESS.

Description of Event or Problem · 1

IT WAS REPORTED BY THE CUSTOMER THAT DURING A PPH PROCEDURE, WHEN THE STAPLER WAS FIRED THE WHITE RING DID NOT BREAK COMPLETELY TO ALLOW STAPLES TO FORM A COMPLETE STAPLE LINE. ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE. NO ADVERSE CONSEQUENCES REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
224517 PROXIMATE PPH PROCEDURE SET STAPLE, IMPLANTABLE GDW ETHICON ENDO-SURGERY, LLC. UNK J4CD55

Patients

Seq Age Sex Outcome Treatment
1