PROXIMATE PPH PROCEDURE SET
Report
- Report Number
- 3005075853-2013-02460
- Event Type
- Malfunction
- Date Received
- May 21, 2013
- Date of Event
- January 11, 2013
- Report Date
- April 29, 2013
- Manufacturer
- ETHICON ENDO-SURGERY, LLC.
- Product Code
- GDW
- PMA / PMN Number
- K051301
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NC, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). BREAKAWAY WASHER CUT OFF CENTER. THE ANALYSIS RESULTS FOUND THAT THE DEVICE ARRIVED WITH THE KNIFE DAMAGED; THE BREAKAWAY WASHER WAS PRESENT AND WITH AN OFF-CENTER CUT. IT APPEARS POSSIBLE THAT THE ANVIL WAS PUSHED FAR ENOUGH OFF CENTER TO RESULT IN AN OFF CENTER CUT OF THE BREAKAWAY WASHER AND DAMAGE THE KNIFE BY PRESSING IT HARD ENOUGH AGAINST THE ANVIL. THIS SITUATION NORMALLY OCCURS WHEN THE TISSUE IS NOT EVENLY DISTRIBUTED IN THE DEVICE. THE DEVICE WAS RELOADED WITH STAPLES, A NEW WASHER WAS PLACED AND THE DEVICE WAS TESTED FOR FUNCTIONALITY, IT FIRED AND FORMED ALL THE STAPLES AS WELL AS COMPLETELY CUT THE TEST MEDIA AND THE BREAKAWAY WASHER WITHOUT INCIDENT, HOWEVER THE WASHER CUT WAS NOT A PERFECT CIRCLE DUE TO THE DAMAGED KNIFE. THE STAPLE LINE WAS COMPLETE AND THE STAPLES WERE NOTED TO HAVE THE PROPER B-FORMED SHAPE. THE MANUFACTURING RECORDS WERE REVIEWED AND NO DISCREPANCIES WERE FOUND DURING THE MANUFACTURING PROCESS.
IT WAS REPORTED BY THE CUSTOMER THAT DURING A PPH PROCEDURE, WHEN THE STAPLER WAS FIRED THE WHITE RING DID NOT BREAK COMPLETELY TO ALLOW STAPLES TO FORM A COMPLETE STAPLE LINE. ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE. NO ADVERSE CONSEQUENCES REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 224517 | PROXIMATE PPH PROCEDURE SET | STAPLE, IMPLANTABLE | GDW | ETHICON ENDO-SURGERY, LLC. | UNK | J4CD55 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |