FDA Adverse Event Injury Summary report: N

HOT SHEARS (TM) (MONOPOLAR CUR

MDR report key: 3122201 · Received May 21, 2013

Report

Report Number
2955842-2013-01764
Event Type
Injury
Date Received
May 21, 2013
Date of Event
March 15, 2012
Report Date
April 26, 2013
Manufacturer
INTUITIVE SURGICAL,INC.
Product Code
NAY
PMA / PMN Number
K050005
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). ON (B)(6) 2013, THE REPORTER CONTACTED ISI AND MENTIONED THAT FRAGMENTS DID NOT FALL INTO THE PATIENT AS INITIALLY INDICATED IN THE INITIAL REPORT THAT WAS SUBMITTED TO THE FDA.

Additional Manufacturer Narrative · 1

PRODUCT WAS REPLACED AND IS BEING REQUESTED BACK FOR INVESTIGATION. THE INSTRUMENT HAS NOT BEEN RETURNED FOR EVALUATION. THE ROOT CAUSE OF THE CUSTOMER REPORTED FAILURE MODE CANNOT BE DETERMINED. A FOLLOW-UP MDR WILL BE SUBMITTED IF THE INSTRUMENT IS RETURNED (POST ENGINEERING EVALUATION) OR IF ADDITIONAL INFORMATION IS RECEIVED. INTUITIVE SURGICAL INC. (ISI) ATTEMPTED TO CONTACT THE REPORTER SEVERAL TIMES VIA EMAIL AND HAS BEEN UNSUCCESSFUL IN GETTING A HOLD OF THE REPORTER TO ANSWER ANY ADDITIONAL QUESTIONS. NO FURTHER CLINICAL INFORMATION WAS PROVIDED. THE COMPLAINT IS BEING REPORTED BASED ON THE INITIAL INFORMATION THAT WAS PROVIDED TO INTUITIVE SURGICAL INC.

Additional Manufacturer Narrative · 1

ON (B)(4) 2013, THE TUBE EXTENSION OF THE MCS INSTRUMENT WAS RETURNED AND EVALUATED. IT WAS NOTED THAT THE TUBE EXTENSION WAS BROKEN AND MISSING A .350 X .275 PIECE AT THE PROXIMAL CLEVIS INTERFACE. THE CLEVIS WAS DISLODGED FROM THE TUBE EXTENSION. THE TUBE EXTENSION WAS FOUND TO BE FRACTURED NEXT TO ONE OF THE KEYS THAT MATES WITH THE PROXIMAL CLEVIS. ENGINEERING EVALUATION CONCLUDED THAT THE DAMAGE WAS LIKELY DUE TO EXCESSIVE SIDE LOADING OR OTHER MISHANDLING/MISUSE. ELECTRICAL CONTINUITY TESTING OF THE INSTRUMENT PASSED. NO OTHER DAMAGE WAS FOUND ON THE DEVICE. THE INSTRUMENTS AND ACCESSORIES USER MANUAL SPECIFICALLY STATES: GENERAL PRECAUTIONS AND WARNINGS O HANDLE INSTRUMENTS WITH CARE. AVOID MECHANICAL SHOCK OR STRESS THAT CAN CAUSE DAMAGE TO THE INSTRUMENTS. DO NOT USE AN INSTRUMENT TO CLEAN DEBRIS FROM ANOTHER INSTRUMENT INTRAOPERATIVELY. THIS MAY RESULT IN DAMAGE TO THE INSTRUMENTS OR OTHER UNINTENDED CONSEQUENCES, SUCH AS DISCONNECTION OF THE INSTRUMENT TIP.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE TIP OF THE MONOPOLAR CURVED SCISSORS BROKE OFF DURING A PROSTATECTOMY PROCEDURE AND FELL INTO THE PATIENT. THE BROKEN TIP WAS RETRIEVED AND THERE WAS NO ALLEGATION OF PATIENT INJURY DUE TO THE ALLEGED ISSUE. THE SURGERY WAS COMPLETED AS PLANNED.