FDA Adverse Event Malfunction Summary report: N

METRX

MDR report key: 3122175 · Received May 21, 2013

Report

Report Number
1030489-2013-01803
Event Type
Malfunction
Date Received
May 21, 2013
Date of Event
March 12, 2013
Report Date
April 29, 2013
Manufacturer
MEDTRONIC SOFAMOR DANEK USA, INC
Product Code
LXH
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CO, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE HAS BEEN RETURNED TO THE MANUFACTURER FOR EVALUATION. VISUALLY CONFIRMED THE INSTRUMENT HANDLE SCREW IS MISSING. OPTICAL INSPECTION OF THE SCREW HOLE AND THE SURROUNDING SURFACE DID NOT REVEAL WITNESS MARKS OR OTHER INDICATIONS REGARDING THE MECHANISM OF FAILURE. THIS INSTRUMENT IS DESIGNED TO BE DISASSEMBLED FOR CLEANING PURPOSES. THE MANUFACTURING LOT CODE SUGGESTS THAT THE INSTRUMENT HAS BEEN IN USE FOR SOME TIME, MAKING IT UNLIKELY THAT THIS IS A MANUFACTURING ISSUE. ADDITIONALLY, THE INFERIOR SHAFT IS BROKEN AT THE CENTERLINE OF THE LOWER PIVOT PIN HOLE. OPTICAL EXAMINATION IDENTIFIED A FAIRLY BRITTLE FRACTURE, WITH NO INDICATION OF FATIGUE. THE LOCATION AND AMOUNT OF FORCE REQUIRED IN ORDER INDUCE FRACTURE OF THE SHAFT IS CONSISTENT WITH OVERLOAD. ALSO THE INSTRUMENT SHAFT IS BENT. REGARDING THE MISSING SCREW NO EVIDENCE WAS FOUND THAT WOULD SUGGEST A DEFECT IN MANUFACTURING OR THE PROCESSING OF THE INSTRUMENT. UNABLE TO DEFINITIVELY DETERMINE SPECIFIC ROOT CAUSE OF THE FOREGOING EVENT FROM THE AVAILABLE INFORMATION. REGARDING THE SHAFT FRACTURE, THE ABOVE OBSERVATIONS ARE CONSISTENT WITH OVERLOAD OF THE INSTRUMENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE BOTTOM JAW BROKE OFF OF THE DOWNBITING PITUITARY AND THE SCREW THAT HOLDS THE HANDLE TOGETHER ALSO FELL OFF DURING AN UNSPECIFIED SPINAL SURGERY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
223683 METRX ORTHOPEDIC MANUAL SURGICAL INSTRUMENT LXH MEDTRONIC SOFAMOR DANEK USA, INC NA GZ04F065

Patients

Seq Age Sex Outcome Treatment
1