FDA Adverse Event
Malfunction
Summary report: N
PROGRAMMING SOFTWARE
MDR report key: 3122162
·
Received May 21, 2013
Report
- Report Number
- 1644487-2013-01500
- Event Type
- Malfunction
- Date Received
- May 21, 2013
- Date of Event
- January 15, 2010
- Report Date
- April 24, 2013
- Manufacturer
- CYBERONICS
- Product Code
- LYJ
- PMA / PMN Number
- P970003
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NC
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
ANALYSIS OF PROGRAMMING HISTORY.
Description of Event or Problem · 1
DURING MANUFACTURER REVIEW OF A PATIENT¿S VNS PROGRAMMING HISTORY, IT WAS IDENTIFIED THAT THE LAST VNS DIAGNOSTICS WERE PERFORMED ON (B)(6) 2010. SETTINGS PRIOR TO THE SYSTEMS TEST WERE 3.25MA/20HZ/250PULSEWIDTH/21SEC ON/0.8MIN OFF. A FINAL INTERROGATION WAS NOT PERFORMED AFTER THE SYSTEMS TEST ON (B)(6) 2010, AND AT THE NEXT OFFICE VISIT ON (B)(6) 2010 THE SETTINGS WERE 0MA/20HZ/500PULSEWIDTH/30SEC ON/60MIN OFF. THIS IS INDICATIVE OF A FAULTED SYSTEMS TEST, EVEN THOUGH THE TEST IS NOT NOTED AS FAULTED IN THE HISTORY. THE VNS SETTINGS WERE CORRECTED ON (B)(6) 2010.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 224772 | PROGRAMMING SOFTWARE | PROGRAMMING SOFTWARE | LYJ | CYBERONICS | MODEL 250 | NI |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 17 YR |