FDA Adverse Event Malfunction Summary report: N

PROGRAMMING SOFTWARE

MDR report key: 3122162 · Received May 21, 2013

Report

Report Number
1644487-2013-01500
Event Type
Malfunction
Date Received
May 21, 2013
Date of Event
January 15, 2010
Report Date
April 24, 2013
Manufacturer
CYBERONICS
Product Code
LYJ
PMA / PMN Number
P970003
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

ANALYSIS OF PROGRAMMING HISTORY.

Description of Event or Problem · 1

DURING MANUFACTURER REVIEW OF A PATIENT¿S VNS PROGRAMMING HISTORY, IT WAS IDENTIFIED THAT THE LAST VNS DIAGNOSTICS WERE PERFORMED ON (B)(6) 2010. SETTINGS PRIOR TO THE SYSTEMS TEST WERE 3.25MA/20HZ/250PULSEWIDTH/21SEC ON/0.8MIN OFF. A FINAL INTERROGATION WAS NOT PERFORMED AFTER THE SYSTEMS TEST ON (B)(6) 2010, AND AT THE NEXT OFFICE VISIT ON (B)(6) 2010 THE SETTINGS WERE 0MA/20HZ/500PULSEWIDTH/30SEC ON/60MIN OFF. THIS IS INDICATIVE OF A FAULTED SYSTEMS TEST, EVEN THOUGH THE TEST IS NOT NOTED AS FAULTED IN THE HISTORY. THE VNS SETTINGS WERE CORRECTED ON (B)(6) 2010.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
224772 PROGRAMMING SOFTWARE PROGRAMMING SOFTWARE LYJ CYBERONICS MODEL 250 NI

Patients

Seq Age Sex Outcome Treatment
1 17 YR