ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2013-06793
- Event Type
- Malfunction
- Date Received
- May 21, 2013
- Report Date
- April 24, 2013
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- KY, US
- Reporter Occupation
- PATIENT
Narratives
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.
DEVICE EVALUATION: THE DEVICE HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(6) 2015 WITH THE FOLLOWING FINDINGS: NO SCRATCHES WERE OBSERVED ON DISPLAY LENS. THE LENS FILM HOWEVER, WAS SCRATCHED. THE DISPLAY SCREEN WAS DIM AND DISCOLORED. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.
ON (B)(6) 2013, THE REPORTER CONTACTED ANIMAS, ALLEGING THAT THE DISPLAY SCREEN LENS WAS DAMAGED. THIS COMPLAINT IS BEING REPORTED BECAUSE THE ALLEGED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE DEVICE CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 224282 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 27 YR |