FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 3121995 · Received May 21, 2013

Report

Report Number
2032227-2013-01896
Event Type
Malfunction
Date Received
May 21, 2013
Date of Event
April 30, 2013
Report Date
April 30, 2013
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE INSULIN PUMP WAS RECEIVED WITH BROKEN DRIVE SUPPORT DISK END CAP. PROTRUDED/LOOSE DRIVE SUPPORT DISK. UNIT RECEIVED WITH CRACKED CASE AT DISPLAY WINDOW CORNERS AND ALARMED DURING REWIND DUE TO CORRODED MOTOR HOME SWITCH.

Description of Event or Problem · 1

CUSTOMER REPORTED THAT THE INSULIN PUMP DRIVE SUPPORT CAP IS STICKING OUT. NOTHING FURTHER WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
224211 PARADIGM REAL-TIME INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC MINIMED MMT-522RNAL

Patients

Seq Age Sex Outcome Treatment
1 46 YR