FDA Adverse Event
Malfunction
Summary report: N
PARADIGM REAL-TIME INSULIN INFUSION PUMP
MDR report key: 3121990
·
Received May 21, 2013
Report
- Report Number
- 2032227-2013-01883
- Event Type
- Malfunction
- Date Received
- May 21, 2013
- Date of Event
- April 28, 2013
- Report Date
- April 29, 2013
- Manufacturer
- MEDTRONIC MINIMED
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- PATIENT
Narratives
Additional Manufacturer Narrative · 1
THE INSULIN PUMP WAS RECEIVED WITH BROKEN DRIVE SUPPORT DISK END CAP.
Description of Event or Problem · 1
CUSTOMER REPORTED THAT SHE HAD HIGH BLOOD GLUCOSE; STATED THAT HER INSULIN PUMP DRIVE SUPPORT CAP IS BROKEN. NOTHING FURTHER WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 223702 | PARADIGM REAL-TIME INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC MINIMED | MMT-722LNAL |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 26 YR |