FDA Adverse Event Injury Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 3121975 · Received May 21, 2013

Report

Report Number
2032227-2013-01889
Event Type
Injury
Date Received
May 21, 2013
Date of Event
April 29, 2013
Report Date
April 29, 2013
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
NE
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE INSULIN PUMP PASSED ALL FUNCTIONAL TEST, INCLUDING PRIME, DISPLACEMENT, REWIND, BASIC OCCLUSION, OCCLUSION, AND EXCESSIVE NO DELIVERY ALARM TEST. UNIT WAS RECEIVED WITH CRACKED DISPLAY WINDOW CORNERS, BATTERY TUBE THREADS AND RESERVOIR TUBE.

Description of Event or Problem · 1

CUSTOMER REPORTED THAT SHE WAS HOSPITALIZED DUE TO HIGH BLOOD GLUCOSE. CUSTOMER'S BLOOD GLUCOSE READING AT THE TIME OF HOSPITALIZATION WAS 522 MG/DL. CUSTOMER STATED THAT SHE TRIED TO BOLUS 25 UNIT, BUT THE INSULIN PUMP DID NOT GIVE IT TO HER. NOTHING FURTHER WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
224972 PARADIGM REAL-TIME INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC MINIMED MMT-722LNAP

Patients

Seq Age Sex Outcome Treatment
1 60 YR Hospitalization