PARADIGM REAL-TIME INSULIN INFUSION PUMP
MDR report key: 3121975
·
Received May 21, 2013
Report
- Report Number
- 2032227-2013-01889
- Event Type
- Injury
- Date Received
- May 21, 2013
- Date of Event
- April 29, 2013
- Report Date
- April 29, 2013
- Manufacturer
- MEDTRONIC MINIMED
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NE
- Reporter Occupation
- PATIENT
Narratives
Additional Manufacturer Narrative · 1
THE INSULIN PUMP PASSED ALL FUNCTIONAL TEST, INCLUDING PRIME, DISPLACEMENT, REWIND, BASIC OCCLUSION, OCCLUSION, AND EXCESSIVE NO DELIVERY ALARM TEST. UNIT WAS RECEIVED WITH CRACKED DISPLAY WINDOW CORNERS, BATTERY TUBE THREADS AND RESERVOIR TUBE.
Description of Event or Problem · 1
CUSTOMER REPORTED THAT SHE WAS HOSPITALIZED DUE TO HIGH BLOOD GLUCOSE. CUSTOMER'S BLOOD GLUCOSE READING AT THE TIME OF HOSPITALIZATION WAS 522 MG/DL. CUSTOMER STATED THAT SHE TRIED TO BOLUS 25 UNIT, BUT THE INSULIN PUMP DID NOT GIVE IT TO HER. NOTHING FURTHER WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 224972 | PARADIGM REAL-TIME INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC MINIMED | MMT-722LNAP |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 60 YR | Hospitalization |