FDA Adverse Event Malfunction Summary report: N

PWRD ECH FLEX 60MM

MDR report key: 3121919 · Received May 21, 2013

Report

Report Number
3005075853-2013-02451
Event Type
Malfunction
Date Received
May 21, 2013
Date of Event
April 11, 2013
Report Date
April 12, 2013
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
GDW
PMA / PMN Number
K110385
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). INCORRECT CARTRIDGE SIZE. ADDITIONAL INFORMATION WAS REQUESTED AND THE FOLLOWING WAS OBTAINED: ON WHAT TISSUE TYPE WAS THE DEVICE USED? COLON OR MESENTERY. AT WHAT LOCATION ON THE TISSUE? ASKU. ON WHICH FIRING(S) DID THIS EVENT OCCUR (1ST, 2ND, 12TH, ETC.)? ASKU. WHAT COLOR CARTRIDGE WAS BEING USED? WHITE. WHAT OTHER COLOR CARTRIDGES WERE USED BEFORE AND AFTER THIS EVENT? ASKU. WAS BUTTRESSING MATERIAL UTILIZED? NO. WAS THE INSTRUMENT FIRED ACROSS OR NEAR AN EXISTING STAPLE LINE OR CLIP? ASKU. WERE ANY UNEXPECTED NOISES HEARD? NO. AFTER USE, DID EACH OF THE TRIGGERS AND BUTTONS AUTOMATICALLY RETURN TO THEIR ORIGINAL (PRE-FIRED) POSITIONS, WITHOUT INTERVENTION? NO, MANUAL OVERRIDE USED TO OPEN DEVICE. WAS THERE ANY DIFFICULTY REMOVING THE DEVICE FROM THE TISSUE? NO. IS THERE ANY OTHER ADDITIONAL INFORMATION YOU WOULD LIKE TO SHARE ABOUT THIS DEVICE OR PROCEDURE? REVERSE BUTTON DID NOT REVERSE KNIFE. WHAT WERE THE PATIENT¿S PRE-EXISTING CONDITIONS? ASKU. DOES THE PATIENT HAVE ANY RELEVANT HISTORY OF SURGICAL TREATMENTS? ASKU. HOW LONG HAS THE SURGEON BEEN USING THIS DEVICE FOR THIS PROCEDURE (IN YEARS)? <1 YEAR. WAS THERE A RECENT CONVERSION TO EES DEVICES IN THIS ACCOUNT OR WITH THIS SURGEON? NO. THE ANALYSIS FOUND THAT ONE PSE60A DEVICE WAS RETURNED IN GOOD VISUAL CONDITION AND WITH THE MANUAL OVERRIDE DOOR OUT OF POSITION; THE OVERRIDE LEVER WAS UP WHICH DENOTES THAT THE KNIFE WAS MANUALLY RETURNED TO HOME POSITION. AN ECR45W CARTRIDGE RELOAD WAS LOADED ON THE DEVICE. THE CARTRIDGE WAS RECEIVED UNFIRED. IT SHOULD BE NOTED THAT A 60MM DEVICE IS DESIGNED TO WORK ONLY WITH 60MM CARTRIDGES, FOR FURTHER LOADING INSTRUCTIONS PLEASE REFER TO THE ECHELON FLEX 60 POWERED IFU. THE DEVICE WAS DISASSEMBLED TO RESET THE BAILOUT SYSTEM AND THEN, THE INSTRUMENT WAS TESTED FOR FUNCTIONALITY IN THE STRAIGHT POSITION WITH A TEST CARTRIDGE RELOAD AND ACHIEVED ITS COMPLETE FIRING SEQUENCE WITHOUT ANY DIFFICULTIES. THE STAPLE LINE AND CUT LINE WERE COMPLETE AND THE STAPLES WERE NOTED TO HAVE THE PROPER B-FORM SHAPE. UPON EVALUATION, THE KNIFE REVERSE BUTTON WAS NOTED FULLY FUNCTIONAL. THE BATCH RECORD WAS REVIEWED AND NO ANOMALIES WERE NOTED DURING THE MANUFACTURING PROCESS.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE DEVICE WAS USED DURING A COLON RESECTION. THE DEVICE MOVED SLIGHTLY BUT THEN STOPPED WORKING AND COULD NOT BE OPENED. NO STAPLES OR CUT LINE WAS ESTABLISHED. BATTERY WAS REMOVED AND REINSERTED, BUT STILL DID NOT WORK. THE REVERSE BUTTON WAS USED BUT DID NOT WORK. THE MANUAL OVERRIDE WAS USED TO RETRACT THE KNIFE BLADE. THE DEVICE WAS OPENED AND REMOVED. THE CASE WAS COMPLETED WITH AN ECHELON DEVICE. THERE WAS NO ADVERSE CONSEQUENCE TO THE PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
224807 PWRD ECH FLEX 60MM STAPLE, IMPLANTABLE GDW ETHICON ENDO-SURGERY, LLC. NA UNK

Patients

Seq Age Sex Outcome Treatment
1 ECR60W