FDA Adverse Event Malfunction Summary report: N

PROGRAMMING SOFTWARE

MDR report key: 3121786 · Received May 21, 2013

Report

Report Number
1644487-2013-01487
Event Type
Malfunction
Date Received
May 21, 2013
Date of Event
January 1, 2013
Report Date
April 25, 2013
Manufacturer
CYBERONICS, INC.
Product Code
LYJ
PMA / PMN Number
P970003
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WI, US
Reporter Occupation
PHYSICIAN

Narratives

Description of Event or Problem · 1

ANALYSIS OF THE COMPUTER AND FLASHCARD WAS COMPLETED. NO ANOMALIES ASSOCIATED WITH THE HANDHELD COMPUTER WERE NOTED DURING TESTING USING THE AC ADAPTER OR THE MAIN BATTERY WITH A FULL CHARGE. THE HANDHELD PERFORMED ACCORDING TO FUNCTIONAL SPECIFICATIONS. DURING THE FLASHCARD ANALYSIS, NO ANOMALIES ASSOCIATED WITH FLASHCARD SOFTWARE OR DATABASES WERE IDENTIFIED. THE FLASHCARD AND SOFTWARE PERFORMED ACCORDING TO FUNCTIONAL SPECIFICATIONS.

Description of Event or Problem · 1

REPORTER INDICATED A VNS DELL X5 COMPUTER WAS FREEZING DURING INTERROGATION AND AT THE ¿INTERROGATION SUCCESSFUL¿ SCREENS. THE SCREEN FREEZING RESOLVED WITH PERFORMING A HARD RESET, BUT THE EVENT RECURS. THE COMPUTER AND FLASHCARD HAVE BEEN RETURNED AND ANALYSIS IS PENDING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
223525 PROGRAMMING SOFTWARE PROGRAMMING SOFTWARE LYJ CYBERONICS, INC. MODEL 250 1073773

Patients

Seq Age Sex Outcome Treatment
1