CODMAN DISPOSABLE PERFORATOR
Report
- Report Number
- 1226348-2013-17397
- Event Type
- Malfunction
- Date Received
- May 21, 2013
- Date of Event
- April 26, 2013
- Manufacturer
- CODMAN & SHURTLEFF, INC.
- Product Code
- HBF
- PMA / PMN Number
- PK791101
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE
Narratives
THE ROOT CAUSE COULD NOT BE DETERMINED. A FUNCTIONAL TEST COULD NOT BE PERFORMED. THE DRIVE PIN FROM THE PERFORATOR ASSEMBLY WAS NOT RETURNED, AND PERFORMING A FUNCTIONAL TEST ON THIS PERFORATOR WITHOUT THE ORIGINAL PIN COULD PRODUCE A DIFFERENT DRILLING RESULT. THE LOT RECORDS HAVE BEEN REVIEWED AND THEY CONFIRMED THAT THE DEVICE CONFORMED TO THE REQUIRED SPECIFICATIONS PRIOR TO DISTRIBUTION.
HISTORICALLY FOR COMPLAINTS OF THIS NATURE, THE RETURNED PERFORATORS HAVE BEEN VISUALLY INSPECTED AS RECEIVED, DISASSEMBLED AND CLEANED, AND THEN VISUALLY INSPECTED. NO DISCREPANCIES HAVE BEEN FOUND. THE PERFORATORS HAVE BEEN REASSEMBLED AND WERE FUNCTIONALLY TESTED FOR CUTTING AND DRILLING. THEY¿VE BEEN FOUND TO MEET SPECIFICATION REQUIREMENTS. THE REPORTED CONDITION COULD NOT BE DUPLICATED. REVIEWS OF THE DEVICE HISTORY RECORDS HAVE FOUND NO DISCREPANCIES. THE COMPLAINTS HAVE NOT BEEN CONFIRMED. A FOLLOW UP REPORT WILL BE FILED IF THE INVESTIGATION OF THIS DEVICE REVEALS A RESULT DIFFERENT FROM THAT WHICH IS STATED ABOVE OR IF ANY FURTHER INFORMATION REGARDING THE CAUSE OR MODE OF FAILURE IS MADE AVAILABLE. OTHERWISE, THE COMPLAINT IS CONSIDERED TO BE CLOSED AT THIS TIME.
THE SALES REP REPORTED THAT THE PERFORATOR WAS INVOLVED IN A PLUNGE WITH A PATIENT. NO FURTHER DETAILS HAVE BEEN REPORTED.